FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 15753487 · Received November 8, 2022

Report

Report Number
1221359-2022-10074
Event Type
Malfunction
Date Received
November 8, 2022
Date of Event
October 19, 2022
Report Date
January 27, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED, SINGLE USE.

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: H10: LOT NUMBERS (202865 AND 210543). LOT NUMBER (202865) EXPIRATION DATE: 07/15/2023 LOT NUMBER (202865) TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 202865 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 202865, TEST BASE PART NUMBER 195-430H / LOT 197969. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 202865 SHOWED THAT THE COMPLAINT RATE IS (B)(4).. LOT NUMBER (210543) TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210543 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210543, TEST BASE PART NUMBER 195-430WL / LOT 207460. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210543 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE FALSE NEGATIVE RESULTS HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. ABBOTT DIAGNOSTICS SCARBOROUGH WILL CONTINUE TO MONITOR AND TREND FOR THIS REPORTED ISSUE. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE.

Additional Manufacturer Narrative · 0

TWO LOT NUMBERS WERE USED BY THE CUSTOMER ON DIFFERENT DATES; HOWEVER, THE CUSTOMER WAS NOT ABLE TO IDENTIFY WHICH LOT NUMBER WAS USED ON SPECIFIC DATES. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6) 2022. THE CONSUMER RECEIVED A NEGATIVE RESULT ON BINAXNOW COVID-19 AG, AND A PCR TEST WAS PERFORMED ON (B)(6) 2022, AND GENERATED A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6)2022. THE CONSUMER RECEIVED A NEGATIVE RESULT ON BINAXNOW COVID-19 AG, AND A PCR TEST WAS PERFORMED ON (B)(6)2022, AND GENERATED A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6) 2022. THE CONSUMER RECEIVED A NEGATIVE RESULT ON BINAXNOW COVID-19 AG, AND A PCR TEST WAS PERFORMED ON (B)(6) 2022, AND GENERATED A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2212885 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 210543 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown