FDA Adverse Event Malfunction Summary report: N

EASYSTAND BANTAM

MDR report key: 1575266 · Received December 9, 2009

Report

Report Number
2183634-2009-00002
Event Type
Malfunction
Date Received
December 9, 2009
Date of Event
November 12, 2009
Report Date
December 2, 2009
Manufacturer
ALTIMATE MEDICAL, INC.
Product Code
INO
PMA / PMN Number
K091242
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DEC 2, 2009. THE ORIGINAL EASYSTAND BANTAM XS WAS RETURNED TO ALTIMATE MEDICAL ON NOV 24, 2009. UPON REVIEW OF THE UNIT, IT WAS CONFIRMED THAT A WELD WAS MISSING ON A BRACKET THAT PROVIDES A MOUNTING POINT FOR THE LINK CONNECTING THE FOOTREST AND KNEE HOLDER ASSEMBLY TO THE BACK BRACKET. THE ABSENCE OF THE WELD ALLOWED PRESSURE FROM THE BACK BRACKET ASSEMBLY, WHICH ALSO CONTAINS THE TRAY MOUNTING ASSEMBLY TO DISPLACE THE BRACKET FROM ITS PROPER POSITION AND EXCEED THE MECHANICAL STRENGTH OF THE BRACKET, THEREBY CAUSING IT BREAK OFF AND ALLOWING THE TRAY/BACK COMBINATION TO FALL FORWARD. THE DESIGN STRENGTH LIES IN THE PRESENCE OF BOTH WELDS DUE TO THE "FORKED" NATURE OF THE BRACKET, IN WHICH IT ENCIRCLES 3 SIDES OF A SQUARE TUBE.

Description of Event or Problem · 1

A CHILD WAS USING AN EASYSTAND BANTAM XS (EXTRA SMALL) WHEN THE BRACKET THAT PROVIDES A MOUNTING POINT FOR THE LINK CONNECTING THE FOOTREST AND KNEE HOLDER ASSEMBLY TO THE BACK BRACKET, FAILED CAUSING THE TRAY/BACK COMBINATION TO FALL FORWARD. THE CHILD WAS NOT INJURED DUE TO THE FACT THAT TWO ATTENDANTS WERE NEARBY TO SUPPORT HER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYSTAND BANTAM 890.3110 INO ALTIMATE MEDICAL, INC. BANTAM XS NA

Patients

Seq Age Sex Outcome Treatment
1 12 YR Other