COOLSCULPTING ELITE SYSTEM
Report
- Report Number
- 3007215625-2022-01679
- Event Type
- Injury
- Date Received
- November 8, 2022
- Date of Event
- March 28, 2022
- Report Date
- May 5, 2025
- Manufacturer
- ALLERGAN PLEASANTON
- Product Code
- OOK
- PMA / PMN Number
- K183514
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ACCORDING TO THE COOLSCULPTING USER MANUAL, UNDER RARE ADVERSE EVENTS, PARADOXICAL ADIPOSE HYPERPLASIA (PH/PAH) IS CHARACTERIZED BY A VISIBLY ENLARGED TISSUE VOLUME WITHIN THE TREATMENT AREA, WHICH MAY DEVELOP TWO TO FIVE MONTHS AFTER TREATMENT. SURGICAL INTERVENTION MAY BE REQUIRED. PAH IS NOT RELATED TO ANY COOLSCULPTING DEVICE FAILURE MODE BUT IT IS OBSERVED MORE FREQUENTLY WITH THE PARALLEL PLATE APPLICATORS. PAH IS INCLUDED IN THE RISK MANAGEMENT FILES OF THE DEVICE AS AN INHERENT RISK TO THE USE OF CRYOLIPOLYSIS FOR LOCALIZED FAT REDUCTION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF AND WHEN ADDITIONAL INFORMATION IS OBTAINED.
CORRECTED DATA: A5 AND D1.
ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED LOWER ABDOMEN AND UPPER BACK WITH COOLSCULPTING ELITE CURVE 120 AND CURVE 150 ON (B)(6) 2022 AND (B)(6) 2022 DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH). DIAGNOSED WITH PAH ON (B)(6) 2022 BY MEDICAL DOCTOR.
ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED LOWER ABDOMEN AND UPPER BACK WITH COOLSCULPTING ELITE CURVE 120 AND CURVE 150 ON (B)(6) 2022 AND (B)(6) 2022 DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH). DIAGNOSED WITH PAH ON (B)(6) 2022 BY MEDICAL DOCTOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2178860 | COOLSCULPTING ELITE SYSTEM | DERMAL COOLING PACK/VACUUM/MASSAGER | OOK | ALLERGAN PLEASANTON | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Female | Required Intervention |