FDA Adverse Event Malfunction Summary report: N

CARELINK WEB UPLOADER

MDR report key: 15751520 · Received November 8, 2022

Report

Report Number
2032227-2022-356977
Event Type
Malfunction
Date Received
November 8, 2022
Date of Event
September 22, 2022
Report Date
November 8, 2023
Manufacturer
MEDTRONIC MINIMED
Product Code
PHV
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION/TESTING SUMMARY: BASED ON ERROR MESSAGE, IT WAS DETERMINED THAT SOME FORM OF ANTI-VIRUS OR SECURITY SOFTWARE WAS INTERFERING WITH UPLOADER APPLICATION. MET W/ CUSTOMER ON A CALL AND USED PROCESS MONITOR APPLICATION TO COLLECT DETAILED LOGS ABOUT WHAT CUSTOMER'S SOFTWARE WAS DOING WHEN TRYING TO USE UPLOADER APPLICATION. WAS ABLE TO IDENTIFY NAMES OF PROCESSES IN LOGS THAT INDICATED THAT CISCO SECURITY SOFTWARE WAS ACTIVE WHILE CUSTOMER WAS USING UPLOADER. (MOST LIKELY) ROOT CAUSE: CISCO SECURITY SOFTWARE INSTALLED ON CUSTOMER'S SYSTEM WAS ATTEMPTING TO PREVENT THE UPLOADER LAUNCH. ANALYSIS SUMMARY: DETERMINED SOURCE OF SOFTWARE INTERFERENCE BASED ON ERROR MESSAGE CUSTOMER WAS RECEIVING AND COLLECTION OF DETAILED DIAGNOSTIC LOGS. PROVIDED CUSTOMER WITH INFORMATION AS WELL AS GUIDANCE ON HOW TO ADD THE NECESSARY EXCLUSIONS AND EXCEPTIONS TO THEIR CISCO SECURITY SOFTWARE. CUSTOMER REPORTED ISSUE AS RESOLVED AFTER APPLYING RECOMMENDED FIXES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER WAS GETTING AN ERROR MESSAGE WITH THE CARELINK PRO AND STATED THAT THE ERROR MESSAGE WAS REGARDING THE SECURITY CHECK OF THE CARELINK UPLOADER APPLICATION THAT FAILED. IT WAS ALSO STATED THAT THE CUSTOMER WAS HAVING ISSUES WITH ADDING THE EXCEPTIONS TO THE AV SOFTWARE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2528597 CARELINK WEB UPLOADER CONTINUOUS GLUCOSE MONITOR RETROSPECTIVE DATA ANALYSIS SOFTWARE PHV MEDTRONIC MINIMED ACC-7350

Patients

Seq Age Sex Outcome Treatment
1 52 YR Unknown