FDA Adverse Event Malfunction Summary report: N

COLON DECOMPRESSION SET

MDR report key: 15751233 · Received November 8, 2022

Report

Report Number
3001845648-2022-00751
Event Type
Malfunction
Date Received
November 8, 2022
Report Date
November 11, 2022
Manufacturer
COOK IRELAND LTD
Product Code
FEG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) # K171619. DEVICE EVALUATION: THE CDSG-14-175 DEVICE OF UNKNOWN LOT NUMBER INVOLVED IN THIS COMPLAINT WAS NOT AVAILABLE FOR EVALUATION. WITH THE INFORMATION PROVIDED, A DOCUMENT-BASED INVESTIGATION WAS CONDUCTED. LAB EVALUATION: THE DEVICE EVALUATION COULD NOT BE COMPLETED AS THE DEVICE OR PHOTOGRAPHIC EVIDENCE OF THE DEVICE WAS NOT RETURNED FOR EVALUATION. DOCUMENT REVIEW: MANUFACTURING RECORDS REVIEW COULD NOT BE COMPLETED AS THE LOT NUMBER IS UNKNOWN. HOWEVER, PRIOR TO DISTRIBUTION ALL CDSG-14-175 DEVICES ARE SUBJECT TO VISUAL A VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. REVIEW OF HISTORICAL DATA: HISTORICAL DATA WAS NOT REVIEWED AS THE LOT NUMBER IS UNKNOWN. THE INSTRUCTIONS FOR USE STATES THE FOLLOWING IN RELATION TO THE INTENDED USE OF THE DEVICE: ¿THE DEVICE IS INTENDED FOR TREATMENT OF ACUTE NON-TOXIC MEGACOLON, PSEUDO-OBSTRUCTION (OGILVIE'S SYNDROME) AND COLONIC STRICTURES." THERE IS EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. IMAGE REVIEW: AN IMAGE WAS NOT RETURNED FOR EVALUATION. ROOT CAUSE REVIEW: A DEFINITIVE ROOT CAUSE OF OFF LABEL USE WAS DETERMINED. IT IS KNOWN THAT THE DEVICE WAS USED IN THE VOLVULUS. THE INSTRUCTIONS FOR USE STATE THAT THE INTENDED USE FOR THE DEVICE IS FOR THE TREATMENT OF ACUTE NON-TOXIC MEGACOLON, PSEUDO-OBSTRUCTION (OGILVIE'S SYNDROME) AND COLONIC STRICTURES. THE USE OF THE CDSG DEVICE IN THE VOLVULUS IS OFF LABEL USE. SUMMARY: COMPLAINT IS CONFIRMED BASED ON CUSTOMER AND/OR REP TESTIMONY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Additional Manufacturer Narrative · 0

PMA/510(K) # K171619. THE INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Description of Event or Problem · 0

SUPPLEMENTAL FOLLOW-UP REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 11-NOV-2022.

Description of Event or Problem · 0

AS REPORTED TO CUSTOMER RELATIONS VIA CRI FORM "THE COLON DECOMPRESSION TUBE WAS DESCRIBED AS ¿TWISTING¿ AT 24 HOURS AFTER PLACEMENT. WHEN QUERIED TO FURTHER DESCRIBE ¿TWISTING¿, THE SITE NOTED THE FOLLOWING IN THEIR QUERY RESPONSE, ¿UNFORTUNATELY, DURING PASSAGE OF THE WIRE, A LARGE LOOP OF WIRE FORMED IN THE DISTENDED VOLVULUS RESULTING IN THE TUBE LOOPING THERE TOO. THE TIP OF THE TUBE IS PROXIMAL TO THE PROXIMAL TWIST THOUGH. THE COLON DECOMPRESSION TUBE WAS REMOVED AT 24 HOURS AFTER PLACEMENT. TREATMENT WITH THE COLON DECOMPRESSION TUBE WAS CONSIDERED SUCCESSFUL AT THE TIME OF REMOVAL AS CONFIRMED BY IMAGING." FINAL PMCF REPORT RECEIVED ON (B)(6) 2022. IT WAS CONFIRMED BY MEDICAL AFFAIRS THAT THE USE OF CDSG DEVICE IN THE VOLVULUS IS OFF LABEL USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2334825 COLON DECOMPRESSION SET FEG TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION FEG COOK IRELAND LTD UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Unknown