UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-00005
- Event Type
- Malfunction
- Date Received
- January 11, 2010
- Date of Event
- December 16, 2009
- Report Date
- April 30, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJC
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
QC BEFORE THE EVENT WAS WITHIN LAB'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE (B)(6)2009. FSE REPLACED THE SAMPLE PROBE AND ALIGNED THE PROBE ARM. NO OTHER ISSUES HAVE BEEN NOTED SINCE SERVICE VISITED. DUE TO A POTENTIAL FOR THIS EVENT TO RECUR UNKNOWINGLY, THERE IS POTENTIAL THAT A PIVOTAL ASSAY COULD BE ERRONEOUSLY REPORTED AND TREATMENT DECISION MAY BE AFFECTED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING FALSELY LOW RESULTS GENERATED BY THE UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. A FALSE LOW TOTAL PROTEIN (TP) RESULT OF 2.1G/DL WAS REPORTED AND QUESTIONED BY THE PHYSICIAN. THE SAMPLE WAS RERUN AND THE RESULT WAS AMENDED TO 6.1G/DL.AN ALBUMIN (ALB) RESULT OF 4.4G/DL WAS REPORTED FOR THIS PATIENT AND UPON REPEAT, THE RESULT WAS AGAIN 4.4G/DL.A GLUCOSE (GLU) RESULT OF 20MG/DL WAS OBTAINED FOR A SECOND PATIENT AND UPON REPEAT, THE RESULT WAS 82MG/DL.THERE WAS NO AFFECT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJC | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |