FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ L CONNECTOR C90J

MDR report key: 15738947 · Received November 4, 2022

Report

Report Number
3003152976-2022-00498
Event Type
Malfunction
Date Received
November 4, 2022
Date of Event
October 12, 2022
Report Date
January 23, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
ONB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 28-NOV-2022. H6: INVESTIGATION SUMMARY: ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. AFTER VISUAL INSPECTION, IT WAS OBSERVED THAT THE MEMBRANE WAS MISSING AND NEVER ASSEMBLED ONTO THE CONNECTOR BODY, THEREFORE THE SYSTEM WAS OPEN AND LEAKED WHEN USED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2201206, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. ASSEMBLY OF THE MEMBRANE IS PERFORMED MANUALLY. A DETECTION SYSTEM IS USED TO VERIFY THE PROPER WELDING AND LOCATION OF THE MEMBRANE. ALL QUALITY RECORDS VERIFY THE INSPECTIONS WERE PERFORMED ACCORDING TO PROCEDURE. WHILE WE CANNOT IDENTIFY A DIRECT ISSUE, IT WAS DETERMINED THIS INCIDENT IS DUE TO THE OPERATOR NOT WELDING THE MEMBRANE ONTO THE CONNECTOR BODY AND THE DETECTION SYSTEM ALSO NOT IDENTIFYING/REJECTING THE IMPACTED PIECE. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT THE BD PHASEAL¿ L CONNECTOR C90J FELL OFF AND DRUG SOLUTION LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER¿S REPORT, THE MEMBRANE OF THE L CONNECTOR FELL OFF AND THE DRUG SOLUTION LEAKED.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT THE BD PHASEAL¿ L CONNECTOR C90J FELL OFF AND DRUG SOLUTION LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER¿S REPORT, THE MEMBRANE OF THE L CONNECTOR FELL OFF AND THE DRUG SOLUTION LEAKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2229285 BD PHASEAL¿ L CONNECTOR C90J INTRAVASCULAR ADMINISTRATION SET ONB BECTON DICKINSON, S.A. 2201206

Patients

Seq Age Sex Outcome Treatment
1 Unknown