FDA Adverse Event Malfunction Summary report: N

DR FOG TREATED SPONGE, STERILE

MDR report key: 15727168 · Received November 4, 2022

Report

Report Number
1836161-2022-00028
Event Type
Malfunction
Date Received
November 4, 2022
Date of Event
October 3, 2022
Report Date
October 3, 2022
Manufacturer
ASPEN SURGICAL PRODUCTS, CALEDONIA
Product Code
OCT
PMA / PMN Number
K932449
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT PRODUCT WAS FOUND WITH SEAL ISSUES. THE ACTUAL DEVICE WAS NOT RETURNED FOR EVALUATION. THE MANUFACTURING LOT NUMBERS WERE PROVIDED FOR REVIEW. PHOTOGRAPHIC EVIDENCE WAS PROVIDED FOR REVIEW AS WELL. THE DISTRIBUTOR INDICATED THAT THE DEFECTS WERE FOUND DURING INCOMING INSPECTION. A REVIEW OF THE PHOTOS CONFIRMED THE ISSUE FROM THE DISTRIBUTOR. ANALYSIS OF THE FINISHED GOOD LOT NUMBERS WERE REVIEWED. NO NON-CONFORMANCE'S WERE NOTED DURING THE MANUFACTURING PROCESS. ACCORDING TO THE MANUFACTURING PROCESS, THESE PARTS ARE MANUALLY LOADED INTO RECESSED POCKETS ON A PACKAGING MACHINE PER SPECIFIED WORK INSTRUCTIONS. IF INCORRECTLY LOADED OR NOT ORIENTATED PROPERLY, THE PRODUCT INTERFERES WITH THE SEALING PROCESS. THEREFORE A LIKELY ROOT CAUSE FOR THE DEFECT MAY BE ATTRIBUTED TO AN OPERATOR ERROR. THE IFU WHICH IS RECEIVED WITH THE PRODUCT, ALONG WITH THE POUCH LABEL, IDENTIFIES THIS FAILURE MODE WITH THE SYMBOL "DO NOT USE IF PACKAGE IS DAMAGED". THIS INDICATES THAT THE DEVICE SHOULD NOT BE USED IF THE PRODUCTS STERILE BARRIER SYSTEM OR ITS PACKAGING IS COMPROMISED. ADDITIONALLY, PRODUCTION SUPERVISORS AND OPERATIONS WERE NOTIFIED OF THIS ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 0

ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT A DR FOG SPONGE WAS DISCOVERED WITH A SEALING ISSUE. THE ITEM WAS NOT IN USE. NO INJURY/DEATH WAS REPORTED. ISSUE WAS REPORTED UNDER THE COMPLAINT HANDLING SYSTEM FOR NUMBER C-1182456.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2783112 DR FOG TREATED SPONGE, STERILE ANTI-FOG OCT ASPEN SURGICAL PRODUCTS, CALEDONIA DF-3120 284279

Patients

Seq Age Sex Outcome Treatment
1 Unknown