COMPANION 5
Report
- Report Number
- 3004972304-2022-00040
- Event Type
- Malfunction
- Date Received
- November 4, 2022
- Report Date
- December 15, 2022
- Manufacturer
- CAIRE INC.
- Product Code
- CAW
- PMA / PMN Number
- K121167
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- 003
Narratives
PURSUANT TO TITLE 21 - FOOD AND DRUGS, CHAPTER I - FOOD AND DRUG ADMINISTRATION DEPARTMENT OF HEALTH AND HUMAN SERVICES, SUBCHAPTER H -0 MEDICAL DEVICE, PART 803 - MEDICAL DEVICE REPORTING, SUBPART A - GENERAL PROVISIONS, SECTION 803.16, NEITHER THIS REPORT NOR ANY INFORMATION SUBMITTED HEREIN CONSTITUTES AN ADMISSION BY CAIRE INC. THAT THE DEVICE STATED IN THIS REPORT, CAIRE INC., OR CAIRE INC.'S EMPLOYEES, CAUSED OR CONTRIBUTED TO THE REPORTABLE EVENT STATED HEREIN. THE DEVICE WAS RETURNED TO CAIRE AND EVALUATED. THE COMPANION 5 UNIT WAS DISPLAYING ALARM CODE AL-P08, INDICATING LOW O2. ALL COMPONENTS INSIDE THE CASE WERE COATED IN DUST. SIEVE DUST WAS OBSERVED BEING EXPELLED FROM THE OUTLET BARB. FOR THE SIEVE DUST TO BE EXPELLED FROM THE BARB, TWO FAULTS MUST HAVE OCCURRED ON THE UNIT. THESE FAULTS ARE SIEVE BED FAILURE AND OUTLET GAS FILTER. FAULT ORDER WAS INCONCLUSIVE, BUT THE UNIT WOULD ALARM FOR THE FAILURE OF THE SIEVE BED.
PURSUANT TO TITLE 21 - FOOD AND DRUGS, CHAPTER I - FOOD AND DRUG ADMINISTRATION DEPARTMENT OF HEALTH AND HUMAN SERVICES, SUBCHAPTER H -0 MEDICAL DEVICE, PART 803 - MEDICAL DEVICE REPORTING, SUBPART A - GENERAL PROVISIONS, SECTION 803.16, NEITHER THIS REPORT NOR ANY INFORMATION SUBMITTED HEREIN CONSTITUTES AN ADMISSION BY CAIRE INC. THAT THE DEVICE STATED IN THIS REPORT, CAIRE INC., OR CAIRE INC.'S EMPLOYEES, CAUSED OR CONTRIBUTED TO THE REPORTABLE EVENT STATED HEREIN. THE DEVICE HAS BEEN RETURNED TO CAIRE FOR AN INVESTIGATION. IF ANY NEW INFORMATION IS DISCOVERED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
THE UNIT WAS RETURNED TO CAIRE REPAIRS WITH OTHER UNITS. THE EVALUATION OF THE DUSTED UNITS DETERMINED THAT THIS UNIT HAD SIEVE DUST COMING THROUGH THE O2 OUTLET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2783880 | COMPANION 5 | CONCENTRATOR, STATIONARY, OXYGEN | CAW | CAIRE INC. | 15067005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |