HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2022-00043
- Event Type
- Malfunction
- Date Received
- November 3, 2022
- Date of Event
- September 29, 2022
- Report Date
- March 27, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002808590
- PMA / PMN Number
- K201658
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CONCLUSION OF CER (B)(4) : HAMILTON MEDICAL AG HAS NOT YET INVESTIGATED THE CASE BASED. HAMILTON MEDICAL IS WAITING FOR THE DEVICE TO INVESTIGATE THE CLAIMED ISSUE " DEVICE WENT IN STANDBY WITHOUT OPERATOR'S INTERACTION" IT WAS CLAIMED AN AMBIENT MODE (VENTILATION STOP), BUT IT IS NOT YET CONFIRMED. NO PATIENT WAS HARMED. THE INVESTIGATION SHOWED THAT A HARDWARE RELATED ISSUE LED TO THE HAMILTON-C6 GOING INTO "AMBIENT STATE". DURING "AMBIENT STATE" THE INSPIRATORY CHANNEL AND EXPIRATORYVALVES ARE OPENED, LETTING THE PATIENT BREATHE ROOM AIR UNASSISTED. CONCLUSION: INVESTIGATION - PENDING. CAUSE: INVESTIGATION - PENDING.
CONCLUSION OF CER (B)(4) : FROM THE INFORMATION RECEIVED AND THE ANALYSIS PERFORMED REGARDING THIS CASE, THE ROOT CAUSE OF THIS EVENT WAS DUE TO A USE ERROR. THE HAMILTON-C6 WORKED AS INTENDED. THE LOG FILE ANALYSIS INDICATE THAT THE VENTILATOR WAS PUT MANUALLY INTO STANDBY MODE. TWO INDEPENDENT STEPS ARE REQUIRED TO GET INTO STANDBY MODE: 1. PRESS THE HARDKEY ON THE INTERACTION PANEL, AND 2. CONFIRM WITH THE TOUCHSCREEN OR 1. PRESS THE HARDKEY ON THE VENTILATOR UNIT, AND 2. CONFIRM WITH THE TOUCHSCREEN. CONCLUSION: THE VENTILATOR WORKED AS INTENDED AND THE USER PUT THE VENTILATIOR INTO STANDBY MODE MANUALLY.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: DESCRIPTION FROM HOSPITAL: AT (B)(6) HOSPITAL THE DEVICE HAS ENTERED AMBIENT STATE WHILE VENTILATING A PATIENT. FIRST INFORMATION HAS IT THAT THE PATIENT WAS NOT INJURED. ORIGINAL CUSTOMER COMPLAINT: THE DEVICE SWITCHED TO AMBIENT MODE DURING VENTILATION.SOME TECHNICAL FAULTS WERE LOGGED.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: ORIGINAL CUSTOMER COMPLAINT: IL CLIENTE DICE CHE IERI ALLE 14.17 IL VENTILATORE È ANDATO IN STAND BY DA SOLO. TRANSLATED WITH WWW.DEEPL.COM: THE CLIENT SAYS THAT YESTERDAY AT 2:17 P.M. THE FAN WENT INTO STANDBY BY ITSELF. *NOTE THE COMPLAINT DESCRIPTION OF THE INTIAL MDR WAS A COPY AND PASTE ERROR AND WAS MISTAKELY OVERTOOK FROM A TEMPLATE. AS WELL THE SERIAL NUMBER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2788963 | HAMILTON MEDICAL AG | HAMILTON-C6 VENTILATOR | CBK | HAMILTON MEDICAL AG | 160021 | 07630002808590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |