CAPTURE-R READY INDICATOR RED CELLS
Report
- Report Number
- 1034569-2010-00005
- Event Type
- Malfunction
- Date Received
- January 8, 2010
- Date of Event
- December 9, 2009
- Report Date
- January 7, 2010
- Manufacturer
- IMMUCOR
- Product Code
- KSZ
- PMA / PMN Number
- BK020053
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
REVIEW OF INSTRUMENT IMAGES: TESTING PERFORMED ON (B)(6) 2009. INTERPRETATION: NEGATIVE. EXPECTED INTERPRETATION: NO PRIOR TESTING. IMAGE DESCRIPTION: LOOKS AS THOUGH THERE IS A WEAK RXN. REACTION STRENGTH: 18. NOTE: CAPTURE-R INDICATOR CELL LOT 221434 EXP 12-03-09 USED DURING THIS TESTING. NEW SAMPLE FROM SAME PATIENT. TESTING PERFORMED ON (B)(6) 2009. INTERPRETATION: POSITIVE. EXPECTED INTERPRETATION: NEGATIVE (BASED ON PREVIOUS TESTING). IMAGE DESCRIPTION: LOOKS POSITIVE. REACTION STRENGTH: 82. NOTE: CAPTURE-R INDICATOR CELL LOT 221441 EXP 12-17-09 USED DURING THESE TEST. REPEAT TESTING; INTERPRETATION: POS. EXPECTED INTERPRETATION: POS (BASED ON PREVIOUS TESTING). IMAGE DESCRIPTION: LOOKS POS. REACTION STRENGTHS: NOT PROVIDED BY THE CUSTOMER (REACTION GRADE 1+). SECOND SAMPLE FROM SAME PATIENT: INTERPRETATION: POS. EXPECTED INTERPRETATION: POS (BASED ON PREVIOUS TESTING). IMAGE DESCRIPTION: LOOKS POS. REACTION STRENGTH: NOT PROVIDED BY THE CUSTOMER (REACTION GRADE 1+). NOTE: IND CELL LOT 22441 EXP 12-17-09 USED DURING TESTING. THE REPEAT TESTING WITH A DIFFERENT LOT OF CAPTURE-R READY INDICATOR CELLS RESULTED AS POSITIVE ON THE SAME INSTRUMENT THAT REPORTED THE INITIAL UNEXPECTED NEGATIVE RESULT. THE INITIAL TESTING PERFORMED ON (B)(6) 2009 SHOWED A SLIGHT REACTION. CANNOT RULE OUT THAT THE VIAL OF CAPTURE-R READY INDICATOR CELLS MAY HAVE BEEN COMPROMISED.
CUSTOMER REPORTED AN UNEXPECTED NEGATIVE REACTION WHEN TESTING A PATIENT SAMPLE BY POOLED_SCREEN ASSAY ON THE GALILEO USING CAPTURE-R READY INDICATOR CELLS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPTURE-R READY INDICATOR RED CELLS | ANTI-IGG COATED INDICATOR RED CELLS FOR USE IN SOLID PHASE ASSAYS | KSZ | IMMUCOR | 221434 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |