T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2022-123387
- Event Type
- Malfunction
- Date Received
- November 2, 2022
- Date of Event
- June 1, 2022
- Report Date
- November 2, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613779
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMIT. PER THE TANDEM USER GUIDE: THE T:SLIM X2 WITH CIQ TECHNOLOGY USER GUIDE STATES: DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. CHANGE YOUR CARTRIDGE EVERY 48¿72 HOURS AS RECOMMENDED BY YOUR HEALTHCARE PROVIDER. WASH YOUR HANDS WITH ANTI-BACTERIAL SOAP BEFORE HANDLING THE INFUSION SET AND THOROUGHLY CLEAN THE INSERTION SITE ON YOUR BODY TO AVOID INFECTION. CONTACT YOUR HEALTHCARE PROVIDER IF YOU HAVE SYMPTOMS OF INFECTION AT YOUR INSULIN INFUSION SITE. THE PUMP USER GUIDE CAUTIONS THAT INSULIN SHOULD BE AT ROOM TEMPERATURE BEFORE USE OR AIR BUBBLES COULD FORM IN THE CARTRIDGE.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. REPORTEDLY, THE CUSTOMER USED ADMELOG INSULIN WITHIN THE CARTRIDGE. ADDITIONALLY, THE CUSTOMER WAS USING COLD INSULIN AND USING CARTRIDGES BEYOND LABELING. TANDEM TECHNICAL SUPPORT EDUCATED CUSTOMER ON PROPER INSULIN/SUPPLY USAGE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 89 MG/DL. REPORTEDLY, THE CUSTOMER CLEARED THE ALARMS AND CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2874160 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613779 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Male | INSULIN: ADMELOG INFUSION SET: AUTOSOFT XC |