ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2022-05187
- Event Type
- Malfunction
- Date Received
- November 2, 2022
- Date of Event
- October 5, 2022
- Report Date
- January 27, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SIMILAR PRODUCT TO 190-000. THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE RELATED MFR. REPORT NUMBERS: 1221359-2022-05188 - 1221359-2022-05193.
INVESTIGATION REPORT: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M213280 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 190-000J/ LOT M213280 AND TEST BASE PART NUMBER: 190-430 / LOT: M213280. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M213280 SHOWED THAT THE COMPLAINT RATE IS (B)(4).. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE.
THE CUSTOMER REPORTED SEVEN FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON MULTIPLE DATES. THIS MFR. REPORT ADDRESSES PATIENT ONE (1) OF SEVEN (7). THE PATIENT TESTED POSITIVE WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2022. CONFIRMATION PCR TESTING GENERATED A NEGATIVE RESULT. NASOPHARYNGEAL SAMPLE WAS USED FOR BOTH THE TESTS. THE PATIENT WAS REPORTEDLY ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2873959 | ID NOW COVID-19 ASSAY | REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | M213280 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |