FDA Adverse Event Injury Summary report: N

MARQUETTE ELECTRONICS, INC.

MDR report key: 157078 · Received March 18, 1998

Report

Report Number
157078
Event Type
Injury
Date Received
March 18, 1998
Date of Event
February 27, 1998
Report Date
March 17, 1998
Manufacturer
MARQUETTE MEDICAL SYSTEMS, INC.
Product Code
MSX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

RADIO TRANSMITTERS USED FOR CARDIAC MONITORING (TELEMETRY MONITORS) WERE OPERATING ON A PREVIOUSLY "UNUSED" TELEVISION BAND. THESE TRANSMITTERS LOST COMMUNICATION WITH THE CENTRAL MONITORING STATION WHEN A NEARBY TELEVISION BROADCAST STATION TURNED ON THEIR NEW HIGH DEFINITION TELEVISION (HDTV OR DTV). IT WAS REVEALED THAT THE HDTV WAS ASSIGNED TO THE SAME CHANNEL BAND. CARDIAC MONITORING COULD NOT BE ACCOMPLISHED WHILE HDTV WAS BROADCASTING WITHIN THE SAME FREQUENCY BAND. THIS OCCURRED ON THE 13TH FLOOR CARDIO-THORACIC SURGERY AND HEART TRANSPLANT NURSING UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MARQUETTE ELECTRONICS, INC. PHYSIOLOGIC MONITOR, TELEMETRY MSX MARQUETTE MEDICAL SYSTEMS, INC. 7700, 9500 NI

Patients

Seq Age Sex Outcome Treatment
1 * Other| R