FDA Adverse Event
Injury
Summary report: N
MARQUETTE ELECTRONICS, INC.
MDR report key: 157078
·
Received March 18, 1998
Report
- Report Number
- 157078
- Event Type
- Injury
- Date Received
- March 18, 1998
- Date of Event
- February 27, 1998
- Report Date
- March 17, 1998
- Manufacturer
- MARQUETTE MEDICAL SYSTEMS, INC.
- Product Code
- MSX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
RADIO TRANSMITTERS USED FOR CARDIAC MONITORING (TELEMETRY MONITORS) WERE OPERATING ON A PREVIOUSLY "UNUSED" TELEVISION BAND. THESE TRANSMITTERS LOST COMMUNICATION WITH THE CENTRAL MONITORING STATION WHEN A NEARBY TELEVISION BROADCAST STATION TURNED ON THEIR NEW HIGH DEFINITION TELEVISION (HDTV OR DTV). IT WAS REVEALED THAT THE HDTV WAS ASSIGNED TO THE SAME CHANNEL BAND. CARDIAC MONITORING COULD NOT BE ACCOMPLISHED WHILE HDTV WAS BROADCASTING WITHIN THE SAME FREQUENCY BAND. THIS OCCURRED ON THE 13TH FLOOR CARDIO-THORACIC SURGERY AND HEART TRANSPLANT NURSING UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MARQUETTE ELECTRONICS, INC. | PHYSIOLOGIC MONITOR, TELEMETRY | MSX | MARQUETTE MEDICAL SYSTEMS, INC. | 7700, 9500 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other| R |