FDA Adverse Event Malfunction Summary report: N

PATROL PUMP

MDR report key: 1570286 · Received December 3, 2009

Report

Report Number
1527460-2009-00138
Event Type
Malfunction
Date Received
December 3, 2009
Date of Event
October 30, 2009
Report Date
November 6, 2009
Manufacturer
ABBOTT NUTRITION
Product Code
LZH
PMA / PMN Number
K943733
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B) (4): THE TESTING AND INVESTIGATION RESULTS WERE RECEIVED AND THE COMPLAINT FOR UNDER DELIVERY WAS CONFIRMED. THE POLES WERE MISALIGNED WITHIN THE MOTOR. (B) (4)

Description of Event or Problem · 1

THE COMPLAINANT REPORTS AN UNDER DELIVERY. THE REPORTER INDICATED THAT THE RATE WAS SET AT 55 ML/HR TO BE DELIVERED OVER 12 HOURS, HOWEVER, THE ACTUAL AMOUNT RECEIVED WAS 37 ML/HR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PATROL PUMP LZH, PUMP, INFUSION, ENTERAL LZH ABBOTT NUTRITION 52036

Patients

Seq Age Sex Outcome Treatment
1 3 YR