FDA Adverse Event
Malfunction
Summary report: N
PATROL PUMP
MDR report key: 1570286
·
Received December 3, 2009
Report
- Report Number
- 1527460-2009-00138
- Event Type
- Malfunction
- Date Received
- December 3, 2009
- Date of Event
- October 30, 2009
- Report Date
- November 6, 2009
- Manufacturer
- ABBOTT NUTRITION
- Product Code
- LZH
- PMA / PMN Number
- K943733
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
(B) (4): THE TESTING AND INVESTIGATION RESULTS WERE RECEIVED AND THE COMPLAINT FOR UNDER DELIVERY WAS CONFIRMED. THE POLES WERE MISALIGNED WITHIN THE MOTOR. (B) (4)
Description of Event or Problem · 1
THE COMPLAINANT REPORTS AN UNDER DELIVERY. THE REPORTER INDICATED THAT THE RATE WAS SET AT 55 ML/HR TO BE DELIVERED OVER 12 HOURS, HOWEVER, THE ACTUAL AMOUNT RECEIVED WAS 37 ML/HR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PATROL PUMP | LZH, PUMP, INFUSION, ENTERAL | LZH | ABBOTT NUTRITION | 52036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |