FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G

MDR report key: 15700069 · Received October 31, 2022

Report

Report Number
1213809-2022-00830
Event Type
Malfunction
Date Received
October 31, 2022
Date of Event
October 5, 2022
Report Date
October 7, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
00382903059164
PMA / PMN Number
K951254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT IS UNKNOWN; AWARENESS DATE HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: THIRTY-EIGHT SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION, NO DEFECTS OR ISSUES WERE OBSERVED. FUNCTIONAL TESTING WAS PERFORMED, EACH SAMPLE WAS CONNECTED TO A SYRINGE WITH SALINE SOLUTION. ALL THE SAMPLES EXPELLED THE SOLUTION WITH NORMAL FLOW, CLOGGED NEEDLE WAS NOT CONFIRMED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2024138. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PROBABLE ROOT CAUSE IS UNKNOWN.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G THE NEEDLE WAS CLOGGED. THIS OCCURRED 5 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER NEEDLE CLOGGED/BLOCKED. CAN YOU CONFIRM THE NUMBER OF OCCURRENCES IN THIS EVENT? MORE THAN 5 TIMES FROM PRODUCT REF 305916 LOT 2024138.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2455789 BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL SYSTEMS 2024138 00382903059164

Patients

Seq Age Sex Outcome Treatment
1 Unknown