BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G
Report
- Report Number
- 1213809-2022-00830
- Event Type
- Malfunction
- Date Received
- October 31, 2022
- Date of Event
- October 5, 2022
- Report Date
- October 7, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMI
- UDI-DI
- 00382903059164
- PMA / PMN Number
- K951254
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT IS UNKNOWN; AWARENESS DATE HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: THIRTY-EIGHT SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION, NO DEFECTS OR ISSUES WERE OBSERVED. FUNCTIONAL TESTING WAS PERFORMED, EACH SAMPLE WAS CONNECTED TO A SYRINGE WITH SALINE SOLUTION. ALL THE SAMPLES EXPELLED THE SOLUTION WITH NORMAL FLOW, CLOGGED NEEDLE WAS NOT CONFIRMED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2024138. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PROBABLE ROOT CAUSE IS UNKNOWN.
IT WAS REPORTED WHILE USING BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G THE NEEDLE WAS CLOGGED. THIS OCCURRED 5 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER NEEDLE CLOGGED/BLOCKED. CAN YOU CONFIRM THE NUMBER OF OCCURRENCES IN THIS EVENT? MORE THAN 5 TIMES FROM PRODUCT REF 305916 LOT 2024138.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2455789 | BD SAFETYGLIDE¿ NEEDLE ONLY, 25 G | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON MEDICAL SYSTEMS | 2024138 | 00382903059164 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |