FDA Adverse Event Malfunction Summary report: N

SDDRIVE SCREWDRIVER T8

MDR report key: 15698282 · Received October 31, 2022

Report

Report Number
8030965-2022-09082
Event Type
Malfunction
Date Received
October 31, 2022
Date of Event
October 5, 2022
Manufacturer
SYNTHES GMBH
Product Code
HXX
UDI-DI
10886982071859
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SDDRIVE SCREWDRIVER T8 WAS FOUND THE TIP STRIPPED. THE OBSERVED CONDITION WAS CONSISTENT AS AN END OF LIFE INDICATOR FOR THE DEVICE. NO OTHER ISSUES WERE IDENTIFIED. THIS CAN BE RELATED WITH THE UNABLE TO ASSEMBLE COMPLAINT. AFTER A VISUAL INSPECTION PER GUIDANCE PROVIDED IN WINDCHILL DOCUMENT, IT WAS DETERMINED THAT THE REUSABLE INSTRUMENT DEVICE WAS FOUND THE TIP STRIPPED FROM REPEATED USE AND SERVICING; THEREFORE, FURTHER INVESTIGATION FOR THE REPORTED COMPLAINT DEVICE IS NOT REQUIRED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED FOUND THE TIP STRIPPED OF THE SDDRIVE SCREWDRIVER T8 WOULD HAVE CONTRIBUTED TO THE COMPLAINED ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. IT WAS DETERMINED THAT THE REUSABLE INSTRUMENT WAS WORN FROM REPEATED USE AND SERVICING. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED? YES, REVIEWED DEVICE HISTORY PART # 03.110.007, SYNTHES LOT # H891253, SUPPLIER LOT # H891253, RELEASE TO WAREHOUSE DATE: 25 SEP 2020, SUPPLIER: (B)(4), NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. INITIAL REPORTER IS A SYNTHES EMPLOYEE. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2022, THE SCREWDRIVER IN QUESTION WAS DROPPED AND AS A RESULT NOW HAS TROUBLE CONNECTING INTO SCREW HEADS. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER T8. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1860823 SDDRIVE SCREWDRIVER T8 SCREWDRIVER HXX SYNTHES GMBH 03.110.007 H891253 10886982071859

Patients

Seq Age Sex Outcome Treatment
1 Unknown