FDA Adverse Event Malfunction Summary report: N

TECNIS SIMPLICITY

MDR report key: 15696815 · Received October 29, 2022

Report

Report Number
3012236936-2022-02718
Event Type
Malfunction
Date Received
October 29, 2022
Date of Event
September 13, 2022
Report Date
December 13, 2022
Manufacturer
AMO PUERTO RICO MFG. INC.
Product Code
HQL
UDI-DI
05050474731776
PMA / PMN Number
P980040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: RETURNED TO MANUFACTURER ON: NOV 7, 2022. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION UNDER MAGNIFICATION REVEALED THAT THE COMPLAINT LENS WAS RECEIVED COATED IN VISCOELASTIC RESIDUE. THE LENS WAS CLEANED AND, SMALL TEARS ON THE HAPTICS COULD BE OBSERVED; ADDITIONALLY, LENS DAMAGE ON THE EDGE OF THE OPTIC BODY COULD BE OBSERVED. A SCRATCH ON THE OPTIC BODY COULD ALSO BE OBSERVED. THE COMPLAINT ISSUE "HAPTIC DAMAGED" WAS OBSERVED DURING PRODUCT EVALUATION; HOWEVER, BASED ON THE COMPLAINT INVESTIGATION RESULTS THE COMPLAINT ISSUE COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS. THE OTHER OBSERVED ISSUES DURING THE PRODUCT EVALUATION COULD NOT BE CONFIRMED TO BE RELATED TO A MANUFACTURING OR DESIGN ISSUE. THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD, COMPLAINT TRENDING, AND RISK DOCUMENTATION FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

THE SURGERY CENTER REPORTED BENT/BROKEN HAPTIC. IT WAS INDICATED THAT THERE WAS NO PATIENT CONTACT WITH THE DEVICE. THERE WAS NO INCISION ENLARGEMENT AND NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2795542 TECNIS SIMPLICITY INTRAOCULAR LENS HQL AMO PUERTO RICO MFG. INC. DIB00 05050474731776

Patients

Seq Age Sex Outcome Treatment
1 Unknown