FDA Adverse Event Malfunction Summary report: N

C.R.E. ESOPHAGEAL/PYLORIC WIREGUIDED BALLOON DILATATION CATHETER

MDR report key: 1569625 · Received December 2, 2009

Report

Report Number
3005099803-2009-05669
Event Type
Malfunction
Date Received
December 2, 2009
Date of Event
November 3, 2009
Report Date
November 3, 2009
Manufacturer
BOSTON SCIENTIFIC CORK, LTD.
Product Code
KNQ
PMA / PMN Number
K971320
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE BALLOON WAS TORN LONGITUDINALLY AND THERE WAS NO DAMAGE TO THE CATHETER SHAFT. THE MOST PROBABLE ROOT CAUSE OF A BALLOON TORN LONGITUDINALLY IS OPERATIONAL CONTEXT. THIS FAILURE OCCURS WHEN ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE LIMITED THE PERFORMANCE OF THE BALLOON (E.G. TORTUOUS ANATOMY). A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO OTHER COMPLAINTS EXIST FOR THE SPECIFIED LOT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORP THAT A C.R.E. ESOPHAGEAL/PYLORIC WIREGUIDED BALLOON DILATATION CATHETER WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY WITH DILATION OF THE ESOPHAGUS. THE PROCEDURE WAS PERFORMED (B)(6) 2009 ON A FEMALE PT. ACCORDING TO THE COMPLAINT, DURING THE PROCEDURE, THE BALLOON WAS POSITIONED WITHIN THE ESOPHAGUS TO TREAT A BENIGN STRICTURE. THE BALLOON WAS SUCCESSFULLY INFLATED WITH SALINE SOLUTION TO 13.5 MM. HOWEVER, WHEN THE PHYSICIAN FURTHER INFLATED THE BALLOON TO 15MM, THE BALLOON BURST. ANOTHER CRE BALLOON OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C.R.E. ESOPHAGEAL/PYLORIC WIREGUIDED BALLOON DILATATION CATHETER KNQ BOSTON SCIENTIFIC CORK, LTD. M00558360 0012842255

Patients

Seq Age Sex Outcome Treatment
1 UNK ALLIANCE HANDLE| SYRINGE