BD¿ ORAL DISPENSING SYRINGE
Report
- Report Number
- 1213809-2022-00829
- Event Type
- Malfunction
- Date Received
- October 29, 2022
- Date of Event
- October 5, 2022
- Report Date
- January 27, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- KYW
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS. MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 16-NOV-2022. H6: INVESTIGATION SUMMARY: ONE SAMPLE AND TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED IS CONFIRMED. POTENTIAL ROOT CAUSES FOR THE MIXED PRODUCT DEFECT CAN BE ASSOCIATED WITH INADEQUATE LINE CLEARANCE, PARTS CONTAINMENT ON THE NEIGHBORING PRODUCTION LINE AND POOR DESIGN OF THE PRINTED BARREL HOPPER ON THE ASSEMBLY MACHINE WHICH POTENTIALLY MIXES FROM THE MATERIAL HANDLING SYSTEM. AS CORRECTIVE ACTIONS, QUALITY ALERT WILL BE ISSUED TO MANUFACTURING FACILITY FOR LINE CLEARANCE ACTIVITY AND ADEQUATE PART CONTAINMENTS. RE-DESIGNING ALSO WILL BE PERFORMED FOR PROTECTIVE COVER ON PRINTED BARREL HOPPER ON ASSEMBLY MACHINE. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1082858. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT A DIFFERENT SYRINGE WAS FOUND IN THE PACK OF BD¿ ORAL DISPENSING SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I WOULD LIKE TO RAISE A SUPPLIER COMPLIANT REGARDING A DIFFERENT MODEL OF SYRINGE FOUND WITHIN AMBER ORAL SYRINGE LOT 1082858."
IT WAS REPORTED THAT A DIFFERENT SYRINGE WAS FOUND IN THE PACK OF BD¿ ORAL DISPENSING SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I WOULD LIKE TO RAISE A SUPPLIER COMPLIANT REGARDING A DIFFERENT MODEL OF SYRINGE FOUND WITHIN AMBER ORAL SYRINGE LOT 1082858."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2691700 | BD¿ ORAL DISPENSING SYRINGE | LIQUID MEDICATION DISPENSER | KYW | BECTON DICKINSON MEDICAL SYSTEMS | 1082858 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |