FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 15692476 · Received October 28, 2022

Report

Report Number
1221359-2022-05195
Event Type
Malfunction
Date Received
October 28, 2022
Date of Event
October 24, 2022
Report Date
February 1, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
04571226475027
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D4 (EXP. DATE) AND H4. TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M213199 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 190-000J/ LOT M213199 AND TEST BASE PART NUMBER 190-430/ LOT M213199. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED, UNCONFIRMED AND CONFLICTING RESULTS) RELATED TO KIT LOT M213199 SHOWED THAT THE COMPLAINT RATE IS (B)(4).. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE.

Additional Manufacturer Narrative · 0

SIMILAR PRODUCT TO 190-000. THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2022. CONFIRMATION PCR TESTING WAS PERFORMED (PLATFORM AND SAMPLE TYPE UNKNOWN) AND GENERATED A NEGATIVE RESULT. THE PATIENT WAS REPORTEDLY ASYMPTOMATIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2920488 ID NOW COVID-19 ASSAY REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M213199 04571226475027

Patients

Seq Age Sex Outcome Treatment
1 Unknown