FDA Adverse Event Malfunction Summary report: N

KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW

MDR report key: 15691054 · Received October 28, 2022

Report

Report Number
3006493760-2022-00002
Event Type
Malfunction
Date Received
October 28, 2022
Date of Event
September 29, 2022
Report Date
October 28, 2022
Manufacturer
ICOTEC AG
Product Code
MQP
UDI-DI
07640164847849
PMA / PMN Number
K200235
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE USER HAS CONFIRMED THAT THE SURGERY WAS PERFORMED IN ACCORDANCE WITH IFU. ANALYSIS OF THE RELEVANT PRODUCTION RECORDS OF THE SCREWDRIVER DID NOT IDENTIFY ANY ANOMALIES THAT COULD BE A CAUSE OF THE EVENT. THE BROKEN SCREWDRIVER WAS RETURNED TO ICOTEC AND THE INVESTIGATION OF THE BREAKAGE HAS BEEN STARTED.

Description of Event or Problem · 0

WHEN TIGHTENING THE LOCKING SCREW OF THE EXPANDABLE VBR, IT WAS NOTICED THAT THE SCREWDRIVER TIP HAD BROKEN OFF. THE BROKEN OFF FRAGMENT REMAINED IN THE HEAD OF THE LOCKING SCREW. THE VBR WAS LEFT IMPLANTED AND THE SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE BROKEN OFF FRAGMENT REMAINED IN THE SCREW HEAD AND THEREFORE IN THE PATIENT'S BODY. THE SUITABILITY OF THE MATERIAL OF THE FRAGMENT (STAINLESS STEEL 1.4112) HAS NOT BEEN EVALUATED FOR LONG-TERM TISSUE/BONE CONTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2725067 KONG-TL VBR SYSTEM, SCREWDRIVER FOR LOCKING SCREW Spinal vertebral body replacement device MQP ICOTEC AG 44-050 20/02 07640164847849

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other