FDA Adverse Event Malfunction Summary report: N

BD PHOENIX PANEL NMIC-306

MDR report key: 15685746 · Received October 27, 2022

Report

Report Number
1119779-2022-01347
Event Type
Malfunction
Date Received
October 27, 2022
Date of Event
October 20, 2022
Report Date
February 3, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
LON
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6. INVESTIGATION SUMMARY: THIS COMPLAINT IS NOT CONFIRMED. THIS COMPLAINT IS FOR FALSE RESISTANCE WITH ERTAPENEM (ETP) AND CEFTAROLINE (CPT) FOR E. COLI (A25922) WHEN USING PHOENIX PANEL NMIC-306 (449292) BATCHES 2196574 AND 2200792. IN ADDITION, THREE (3) RETENTION PANELS FROM THE COMPLAINT BATCH 200792 WERE INOCULATED WITH QC ISOLATE A25922, AND THREE (3) RETENTION PANELS FROM COMPLAINT BATCH 2196754 WERE INOCULATED WITH QC ISOLATE 14780 TO EVALUATE FOR MIC RESULTS. ALL PANELS RETURNED SATISFACTORY RESULTS, THEREFORE THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS REVEALED NO QUALITY NOTIFICATIONS GENERATED ON THE COMPLAINT BATCH. A REVIEW OF COMPLAINTS REVEALED NO ADDITIONAL COMPLAINTS ON THE COMPLAINT BATCH. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD ID/AST PLANT QUALITY WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS. THE FOLLOWING ARE GUIDELINES TO HELP MINIMIZE PHOENIX AST ISSUES: Ø MEDIA SELECTION O ISOLATES MUST BE RECOVERED FROM NON-SELECTIVE MEDIA. SEE TABLE 16 RECOMMENDED MEDIA IN THE BD USER MANUAL FOR A LIST OF RECOMMENDED MEDIA. O ENSURE QUALITY OF VENDOR SELECTION FOR PLATED MEDIA. VARIATIONS IN FORMULATIONS MAY IMPACT RESULTS. Ø CULTURE HANDLING O CULTURES MUST BE 18-24 HOURS OLD FOR GRAM-NEGATIVE & GRAM-POSITIVE ORGANISMS AND 18-48 HOURS OLD FOR YEAST ORGANISMS. O FOR QC ORGANISMS, ENSURE ORGANISMS HAVE BEEN SUBCULTURED AT LEAST TWICE ON TWO CONSECUTIVE DAYS ON PROPER NONSELECTIVE MEDIA (GRAM-NEGATIVE OR GRAM-POSITIVE ORGANISMS: TSA WITH 5% SHEEP BLOOD, YEAST: SABOURAUD DEXTROSE AGAR). Ø MCFARLAND PREPARATION O USE OF A LOW-QUALITY STERILE COTTON SWAB, WHICH SHED FIBERS, COULD POTENTIALLY CONTRIBUTE TO A FALSELY HIGH MCFARLAND READING. POLYESTER SWABS ARE NOT RECOMMENDED. O ENSURE BD APPROVED NEPHELOMETER IS ADEQUATELY CALIBRATED WITH NOT EXPIRED MCFARLAND CALIBRATION TUBES. O PREPARED TUBES SHOULD BE VORTEXED FOR 5 SECONDS AND ALLOWED APPROXIMATELY 10 SECONDS FOR AIR BUBBLES TO SURFACE. O ENSURE MCFARLAND FALLS WITHIN PROPER RANGE OF THE INOCULUM DENSITY. FOR 0.5 MCFARLAND SYSTEM, 0.50-0.60 IS ACCEPTABLE. FOR 0.25 MCFARLAND SYSTEM, 0.20-0.30 IS ACCEPTABLE. FOR YEAST PANELS, 2.00-2.40 IS ACCEPTABLE. O CONFIRM CURRENT INSTRUMENT SETTINGS FOR INOCULUM DENSITY BEFORE INOCULATING PANELS. FOR EXAMPLE, ENSURE A 0.50 MCFARLAND WAS NOT PREPARED FOR A 0.25 INOCULUM DENSITY INSTRUMENT SETTING. USE BACTERIAL SUSPENSIONS WITHIN 60 MINUTES OF PREPARATION. Ø PANEL PREPARATION PANELS SHOULD BE USED WITHIN 2 HOURS OF REMOVAL FROM POUCH. INOCULATE PANELS WITHIN 30 MINUTES OF THE TIME THAT THE AST BROTH INOCULUM IS PREPARED. ALLOW SUFFICIENT TIME FOR FLUID TO TRAVERSE DOWN THE WELL TRACKS BEFORE MOVING THE PANEL. AVOID TOUCHING THE FRONT AND BACKSIDE OF THE PANEL. HANDLE PANELS BY THE SIDES TO AVOID PRODUCING SMUDGES ON THE SURFACE OF THE PANELS. INOCULATED PANELS SHOULD BE HANDLED WITH CARE. AVOID KNOCKING OR JARRING THE PANEL. LOAD PANELS INTO INSTRUMENT WITHIN 30 MINUTES OF INOCULATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PHOENIX PANEL NMIC-306 IS FLAGGING RESISTANT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER STATES PANEL IS FLAGGING RESISTANT BUT WHEN PLATED IT IS SENSITIVE HAZARD, INJURY OR ERRONEOUS RESULTS? NO. HAZARD, INJURY OR ERRONEOUS RESULTS DETAILS ".

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PHOENIX PANEL NMIC-306 IS FLAGGING RESISTANT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER STATES PANEL IS FLAGGING RESISTANT BUT WHEN PLATED IT IS SENSITIVE HAZARD, INJURY OR ERRONEOUS RESULTS? NO HAZARD, INJURY OR ERRONEOUS RESULTS DETAILS "

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2480370 BD PHOENIX PANEL NMIC-306 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY LON BECTON, DICKINSON & CO. (SPARKS) 2200792

Patients

Seq Age Sex Outcome Treatment
1 Unknown