FDA Adverse Event Injury Summary report: N

OCTRODE LEAD KIT, 60CM LENGTH

MDR report key: 15682744 · Received October 27, 2022

Report

Report Number
3006705815-2022-17675
Event Type
Injury
Date Received
October 27, 2022
Date of Event
October 3, 2022
Report Date
December 14, 2022
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
Product Code
LGW
UDI-DI
05415067017246
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING PROCESSING OF THIS INCIDENT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE PATIENT INFORMATION.

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE DOCUMENTS REVIEWED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 0

RELATED MANUFACTURER REFERENCE NUMBER:3006705815-2022-17676. IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION OBTAINED INDICATES THE LEADS WERE REPLACED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2803036 OCTRODE LEAD KIT, 60CM LENGTH PERCUTANEOUS LEAD LGW ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) 3186 A000103584 05415067017246

Patients

Seq Age Sex Outcome Treatment
1 43 YR Female Other SCS IPG.