BINAXNOW COVID-19 AG SELF TEST 2CT
Report
- Report Number
- 1221359-2022-10032
- Event Type
- Malfunction
- Date Received
- October 27, 2022
- Date of Event
- October 4, 2022
- Report Date
- January 25, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210526 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210526, TEST BASE PART NUMBER 195-430H / LOT 207568. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND/OR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210526 SHOWED THAT THE COMPLAINT RATE IS (B)(4).. ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED, SINGLE USE.
THE CONSUMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6)2022. THE CONSUMER PERFORMED TWO TESTS ON(B)(6)2022, THE FIRST TEST WAS A POSITIVE RESULT, AND THE SECOND TEAT WAS A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
THE CONSUMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6) 2022. THE CONSUMER PERFORMED TWO TESTS ON (B)(6) 2022, THE FIRST TEST WAS A POSITIVE RESULT, AND THE SECOND "TEST" WAS A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2572587 | BINAXNOW COVID-19 AG SELF TEST 2CT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 210526 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |