FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 15678075
·
Received October 26, 2022
Report
- Report Number
- 9610877-2022-59272
- Event Type
- Malfunction
- Date Received
- October 26, 2022
- Date of Event
- October 4, 2022
- Report Date
- October 27, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- GCL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EE-1580K IS AVAILABLE IN THE USA WITH A 510K NUMBER K961564. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2021442 | PENTAX | VIDEO ESOPHAGOSCOPE | GCL | HOYA CORPORATION PENTAX TOKYO OFFICE | EE17-J10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |