PLATE TRYPTICASE SOY AGAR 5% SB 100 EA
Report
- Report Number
- 1119779-2022-01333
- Event Type
- Malfunction
- Date Received
- October 26, 2022
- Date of Event
- October 18, 2022
- Report Date
- October 31, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JSG
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY: COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINT HAS BEEN TAKEN ON THIS LOT. AT THIS TIME. DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY AND NO DEVIATIONS WERE OBSERVED. IT INCLUDES BIOBURDEN TESTING. A VISUAL INSPECTION WAS PERFORMED ON 20 RETENTION SAMPLES OF THIS LOT. AS A RESULT,NO CONTAMINATION WAS OBSERVED. THE ROOT CAUSE FOR THIS COULD NOT BE DETERMINED WE WILL CONTINUE TO MONITOR THIS LOT.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING PLATE TRYPTICASE SOY AGAR 5% SB 100 EA CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, BACILLUS LICHENIFORMIS WAS DETECTED IN ONE PLATE OUT OF 1000 (100 PLATES X 10 CARTONS) BEFORE USAGE.".
IT WAS REPORTED THAT WHILE USING PLATE TRYPTICASE SOY AGAR 5% SB 100 EA CONTAMINATION OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, BACILLUS LICHENIFORMIS WAS DETECTED IN ONE PLATE OUT OF 1000 (100 PLATES X 10 CARTONS) BEFORE USAGE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2802917 | PLATE TRYPTICASE SOY AGAR 5% SB 100 EA | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON, DICKINSON & CO. (SPARKS) | 2228492 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |