FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 15670995 · Received October 25, 2022

Report

Report Number
1911916-2022-00587
Event Type
Malfunction
Date Received
October 25, 2022
Date of Event
October 1, 2022
Report Date
December 16, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
00382903096282
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 07-NOV-2022. H6: INVESTIGATION SUMMARY: ONE SAMPLE AND TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION, A BROWNISH DISCOLORATION WAS OBSERVED ACROSS THE BARREL FLANGES EXTENDING DOWN THE NECK OF THE BARREL. POTENTIAL ROOT CAUSE FOR THE EMBEDDED FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. THE EMBEDDED FOREIGN MATTER IS MOST LIKELY DEGRADED PLASTIC. THIS OCCURS WHEN THE RESIN IS EXPOSED TO PROLONGED HIGH TEMPERATURES INSIDE THE MOLDING MACHINE, SUCH AS DURING START UP. THIS TYPE OF DEFECT IS COSMETIC AND DOES NOT POSE RISK TO THE CUSTOMER. THESE CONDITIONS ARE OCCURRING BELOW THEIR EXPECTED FREQUENCY SO NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2042709. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE EXPERIENCED A DISCOLORATION OF SOLUTION ONCE DRAWN INTO THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN DRAWING UP RADIOPHARMACEUTICALS IN THE 1ML SYRINGES, A BROWN DISCOLORATION WAS NOTICED. NO SERIOUS CONSEQUENCES FOR THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE EXPERIENCED A DISCOLORATION OF SOLUTION ONCE DRAWN INTO THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN DRAWING UP RADIOPHARMACEUTICALS IN THE 1ML SYRINGES, A BROWN DISCOLORATION WAS NOTICED. NO SERIOUS CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2760353 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 2042709 00382903096282

Patients

Seq Age Sex Outcome Treatment
1 Unknown