FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 15666264 · Received October 25, 2022

Report

Report Number
1221359-2022-10025
Event Type
Malfunction
Date Received
October 25, 2022
Date of Event
October 1, 2022
Report Date
January 25, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED, SINGLE USE.

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: D4: LOT NUMBER TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 211329 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 211329, TEST BASE PART NUMBER 195-WJR / LOT 207739. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND/OR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT. 211329 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SPECIFIC PATIENT SAMPLE. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF TEST ON (B)(6) 2022. THE CUSTOMER GOT A NEGATIVE RESULT AFTER GETTING A POSITIVE RESULT WHEN FIRST TEST WAS PERFORMED. ALSO, THE CUSTOMER STATED GETTING A NEGATIVE RESULT TWO DAYS AGO. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF TEST ON (B)(6)2022. THE CUSTOMER GOT A NEGATIVE RESULT AFTER GETTING A POSITIVE RESULT WHEN FIRST TEST WAS PERFORMED. ALSO, THE CUSTOMER STATED GETTING A NEGATIVE RESULT TWO DAYS AGO. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2629425 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 211329 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 26 YR Male