FDA Adverse Event Malfunction Summary report: N

BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM

MDR report key: 15663993 · Received October 24, 2022

Report

Report Number
3014704491-2022-00512
Event Type
Malfunction
Date Received
October 24, 2022
Date of Event
August 28, 2022
Report Date
October 20, 2022
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903839124
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER E-MAIL: UNKNOWN. INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED, A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1221451. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM LEAKED FROM THE PRN DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "AT 15:30 ON (B)(6) 2022, INTRAVENOUS FLUID WAS GIVEN TO THE PATIENT, AND AN INDWELLING VENOUS INDWELLING NEEDLE WAS USED FOR THE PATIENT. AFTER EMPTYING THE AIR, THE PRN JOINT OF THE INDWELLING NEEDLE WAS FOUND TO BE LEAKING, AND A NEW INDWELLING NEEDLE WAS IMMEDIATELY REPLACED, WHICH CAN BE USED NORMALLY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340940 BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 1221451 00382903839124

Patients

Seq Age Sex Outcome Treatment
1 Unknown