FDA Adverse Event Other Summary report: N

BAIR PAWS

MDR report key: 1566374 · Received December 8, 2009

Report

Report Number
3004542876-2009-00004
Event Type
Other
Date Received
December 8, 2009
Date of Event
December 19, 2008
Report Date
December 7, 2009
Manufacturer
ARIZANT HEALTHCARE INC.
Product Code
DWJ
Removal / Correction Number
NOT YET ASSIGNED
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALTHOUGH THE ADVERSE EVENT OCCURRED ON (B)(6) 2008, WE DID NOT BECOME AWARE OF INFO THAT WOULD LEAD US TO CONCLUDE THAT IT WAS A REPORTABLE EVENT UNTIL NOVEMBER 19, 2009. OUR INITIAL INVESTIGATION FOUND THE EVENT WAS DUE TO UNUSUALLY ABUSIVE USE AND HANDLING AND WAS NOT A DEVICE MALFUNCTION. ADD'L MATERIAL INFO HAS MORE RECENTLY BEEN ACQUIRED FROM INDUSTRY COMMUNICATIONS AND RECALL NOTICES CONCERNING DEFECTIVE AC POWER CORDS MANUFACTURED BY ELECTRI-CORD MFG CORP THAT ALTERED THIS INITIAL CONCLUSION.

Description of Event or Problem · 1

REPORT OF ELECTRI-CORD POWER CORD SMOKING AT WALL END PLUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BAIR PAWS PATIENT WARMING THERAPY DWJ ARIZANT HEALTHCARE INC. 850 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other