FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 15659091 · Received October 24, 2022

Report

Report Number
3004464228-2022-19875
Event Type
Malfunction
Date Received
October 24, 2022
Date of Event
October 19, 2022
Report Date
October 20, 2022
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION TO D(4): CATALOG NO CHANGED FROM UNAVAILABLE TO ZXP425. CORRECTION TO G(5): PMA/510(K) # CHANGED FROM UNAVAILABLE TO K192659.

Additional Manufacturer Narrative · 0

ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR DETERMINE ITS ROOT CAUSE. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.

Description of Event or Problem · 0

IT WAS REPORTED THE CANNULA DEPLOYED EARLY, INDICATING A FAILURE OF THE NEEDLE MECHANISM. THE PATIENT'S BLOOD GLUCOSE LEVELS WERE UNAFFECTED, AND THE POD WAS NOT WORN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2060128 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 40160 L50364

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female