FDA Adverse Event Malfunction Summary report: N

630G INSULIN PUMP MMT-1715KL 630G

MDR report key: 15656973 · Received October 24, 2022

Report

Report Number
2032227-2022-351722
Event Type
Malfunction
Date Received
October 24, 2022
Date of Event
October 4, 2022
Report Date
February 22, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000316655
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

S.W VERSION 4.11C. RETAINER RING = BLACK. CASE TYPE = NGP. CUSTOMER RETURNED PUMP FOR AN ALLEGED NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM AND COSMETIC DAMAGE LOCATED AT THE BELT CLIP RAILS FOUND ON 04-OCT-2022. THE PUMP WAS RECEIVED WITH A BROKEN BELT CLIP RAILS. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08715 INCHES. THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO UNEXPECTED OCCLUSIONS OR INSULIN FLOW BLOCKED ALARM NOTED DURING TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. THERE WAS ¿NO¿ NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM RECORDED IN THE FORMATTED HISTORY FILE FOR THE EVENT DATE. HOWEVER, NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM WAS FOUND IN THE FORMATTED HISTORY FILE ON: 09/06/2022 14:25:56.000, 09/06/2022 14:27:01.000, 09/15/2022 07:54:16.000. 09/17/2022 17:50:45.000, 09/17/2022 17:52:39.000, 09/17/2022 17:53:09.000. AND 09/20/2022 12:46:51.000 DURING BOLUS DELIVERY. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA1, PCBA2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. FORCE SENSOR ZERO OFFSET WITHIN SPECIFICATION (24.0 MV). THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY AND A SCRATCHED CASE. COSMETIC DAMAGE WAS CONFIRMED AT THE BELT CLIP RAILS. NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM WAS NOT CONFIRMED. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

THE DEVICE WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAD REPORTED THAT THE PUMP NO LONGER HOLDS THE CLIP AND THE PUMP WAS ALARMING INSULIN FLOW BLOCK. CUSTOMER STATED THAT THEY HAVE BEEN EXCHANGING THE INFUSION PORT AND CONTINUING WITH THE SAME ISSUE. THE TROUBLESHOOTING WAS NOT PERFORMED AS IT WAS DECLINED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR PRODUCT ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2149645 630G INSULIN PUMP MMT-1715KL 630G ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715KL HG4EBN6 000000763000316655

Patients

Seq Age Sex Outcome Treatment
1 58 YR Female