FDA Adverse Event Injury Summary report: N

VASERLIPO SYSTEM

MDR report key: 15654884 · Received October 21, 2022

Report

Report Number
3011423170-2022-00132
Event Type
Injury
Date Received
October 21, 2022
Date of Event
July 24, 2022
Report Date
September 23, 2022
Manufacturer
SOLTA MEDICAL, INC
Product Code
QPB
UDI-DI
00850608002537
PMA / PMN Number
K190551
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED FOR AN EVALUATION. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

CORRECTION: SECTION D2 PRODUCT CODE FROM: MUU - TO: QPB, SECTION G4: FROM: K110306 - TO: K190551. THE CUSTOMER HAS BEEN ADVISED TO STOP USING THE DEVICE. ATTEMPTS HAVE BEEN MADE TO GET THE DEVICE BACK. IF THE DEVICE IS RETURNED AND EVALUATED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

NO OTHER INFORMATION WAS PROVIDED BY THE CUSTOMER. NO PRODUCT WAS RETURNED FOR EVALUATION AFTER MULTIPLE FOLLOW UP ATTEMPTS WERE MADE. THE SAFE AND EFFECTIVE USE OF THIS MEDICAL DEVICE DEPENDS TO A LARGE DEGREE ON FACTORS SOLELY UNDER THE CONTROL OF THE OPERATOR. IT IS IMPORTANT THAT ALL OPERATORS OF THE VASER SURGICAL SYSTEM READ, UNDERSTAND, AND FOLLOW THE OPERATING INSTRUCTIONS SUPPLIED WITH EQUIPMENT. USE OF THIS DEVICE IS LIMITED TO THOSE PHYSICIANS WHO, BY MEANS OF FORMAL PROFESSIONAL TRAINING OR SANCTIONED CONTINUING MEDICAL EDUCATION (INCLUDING SUPERVISED OPERATIVE EXPERIENCE), HAVE ATTAINED PROFICIENCY IN SUCTION LIPOPLASTY. A REVIEW OF MANUFACTURING RECORDS SHOWED ALL REQUIREMENTS WERE MET. FINAL TEST VERIFICATION SPECIFICATIONS ARE ACCEPTABLE. NO NON-CONFORMITIES OR ANOMALIES FOUND RELATED TO THIS COMPLAINT WHEN REVIEWING THE DEVICE HISTORY RECORD. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. DUE TO LACK OF INFORMATION, NO CAUSAL FACTORS CAN BE DETERMINED, AND NO CONCLUSIONS CAN BE DRAWN. AS A RESULT, THE ROOT CAUSE IS UNKNOWN. NO CORRECTIVE ACTION IS REQUIRED.

Additional Manufacturer Narrative · 0

CORRECTION: B3: DATE OF EVENT - FROM 22JUL2022 TO 24JUL2022. TREATMENT SETTINGS WERE APPROPRIATE. IT WAS REPORTED LOW AMPLITUDE POWER WAS USED. THE AMOUNT OF VASER TIME WAS NORMAL FOR THE PROCEDURE. A LARGE AND SAFE AMOUNT OF WETTING SOLUTION WAS INFILTRATED. THE DEVICE AND HANDPIECES WERE RETURNED FOR EVALUATION. VASER AMPLIFIERS ARE EQUIPPED WITH AUDIBLE AND VISUAL INDICATORS (ALERT/WARNING) FOR SYSTEM FAULT CONDITIONS. THE AUDIBLE ALERT/WARNING SIGNAL INDICATOR IS AN URGENT WARNING TONE WHICH CAN OCCUR IF THE SYSTEM DETECTS CONDITIONS OUTSIDE OF SAFETY PARAMETERS PRIOR TO AND DURING TREATMENT. NO DEVICE OR HAND PIECE ISSUES WERE FOUND DURING THE EVALUATION. THE VASER AMPLIFIER PASSED ALL INCOMING CHECKS, SELF-TESTS, AND ULTRASOUND BOARD TESTING. THE HAND PIECES ALSO PASSED FUNCTIONAL AND SAFETY TESTING. AN INDEPENDENT MEDICAL REVIEW WAS PERFORMED BY A BOARD-CERTIFIED PHYSICIAN WITH EXTENSIVE VASER EXPERIENCE ON THE INFORMATION AVAILABLE REGARDING THIS EVENT. THE REVIEW OF THIS EVENT FOUND IT DOES NOT APPEAR THAT THE PATIENT BURNS WERE ATTRIBUTABLE TO THE VASER AMPLIFIER. BASED ON THE MEDICAL REVIEW, THIS EVENT REPRESENTS TECHNICAL MISTAKES ON THE PART OF THE SURGEON AND HAS NO RELATIONSHIP TO A MALFUNCTION OF THE VASER DEVICE. A REPRESENTATIVE FOR SOLTA MEDICAL VISITED THE DOCTOR AFTER THIS EVENT. IT WAS VERIFIED THAT THE PHYSICIAN USED THE VASER SYSTEM CORRECTLY AND HAD PERFORMED THE PROCESS SUCCESSFULLY. THE PHYSICIAN REPORTED THAT THEY ARE NOW USING A DIFFERENT STYLE OF SOLTA PROBE AND DEVICE SETTING AFTER THESE EVENTS OCCURRED. NO ADVERSE EVENTS HAVE OCCURRED SINCE THE PHYSICIAN MADE THESE CHANGES. IT WAS DETERMINED NO ADDITIONAL TRAINING WAS REQUIRED. A REVIEW OF THE DEVICE HISTORY RECORD FOR THE VASER AMPLIFIER THAT WAS IN USE AT THIS CLINIC (SN BBD34Y), SHOWED THE SYSTEM PASSED ALL REQUIRED FUNCTIONAL AND FINAL TESTING DURING MANUFACTURING. NO NON-CONFORMITIES OR ANOMALIES FOUND RELATED TO THIS COMPLAINT WHEN REVIEWING THE DEVICE HISTORY RECORD. SERVICE RECORDS SHOW THAT THERE HAVE BEEN NO COMPLAINTS OR EVENTS REPORTED FOR THIS VASER AMPLIFIER PRIOR TO THESE EVENTS. POST-MARKET ANALYSIS SHOWS THE RISK OF PATIENT BURNS FROM A LIPOSUCTION PROCEDURE WHEN USING VASER IS IMPROBABLE. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. BASED ON THE MEDICAL REVIEW, THIS EVENT REPRESENTS TECHNICAL MISTAKES ON THE PART OF THE SURGEON AND HAS NO RELATIONSHIP TO A MALFUNCTION OF THE VASER DEVICE. NO CORRECTIVE ACTION IS REQUIRED.

Description of Event or Problem · 0

A USER FACILITY REPORTED A BURN FOLLOWING SURGERY WITH A VASER PRO AMPLIFIER. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED BY THE USER FACILITY: A USER FACILITY REPORTED BURNS TO THE PATIENT'S ABDOMEN AND FLANKS TWO DAYS POST OPERATION FROM A VASERLIPO PROCEDURE. SECONDARY INTERVENTION (OINTMENTS, MEDICATIONS, ETC.) INVOLVED BURN CREAMS. IT IS NOTED THAT THERE WILL BE PERMANENT DAMAGE OR SCARRING. AVAILABLE PATIENT PICTURES WERE REVIEWED BY THE MEDICAL EXAMINER. A LARGE, BURNED AREA IS VISIBLE ON LOWER ABDOMEN ON BOTH SIDES EXTENDING TO THE FLANKS. IN ONE PICTURE OPEN WOUND IS VISIBLE ON ONE SIDE OF LOWER ABDOMEN. LARGE AREA OF POST INFLAMMATORY HYPERPIGMENTATION IS VISIBLE ON LOWER AND MIDDLE ABDOMEN AREAS. NO OTHER TREATMENTS (BESIDES VASER) WERE BEING PERFORMED IN SAME AREA WHERE SYMPTOMS WERE REPORTED. THE PATIENT HASN'T UNDERGONE ANY OTHER TREATMENTS IN THE SAME SYMPTOM AREA WITHIN THE PAST 90 DAYS. THE VASER SELF-TEST WAS PERFORMED PRIOR TO THE PROCEDURE. THE PATIENT WAS ALREADY UNDER ANESTHESIA WHEN THE SYSTEM WAS TESTED. THE SYSTEM WAS IN VASER (V) MODE. THE HIGHEST AMPLITUDE LEVEL USED WAS 30% WHERE THE INJURY HAPPENED. NO SYSTEM ERRORS OCCURRED NOR WAS ANYTHING OUT OF THE ORDINARY NOTICED DURING TREATMENT. THE TOTAL TIME OF THE VASER DELIVERY WAS 6 MINUTES. APPROXIMATELY 6000-8000 ML OF TUMESCENT FLUID WAS USED. THE PROCEDURE WAS COMPLETED USING THE VASER. A SKIN PORT WAS USED AND WAS NOT DAMAGED OR MISALIGNED. A WET TOWEL BARRIER WAS UTILIZED TO PROTECT THE SKIN FROM PROBE CONTACT. A 1 AND 2 RING PROBE WAS USED FOR THE TREATMENT. THIS IS NOT THE FIRST TIME THE PROBE/CANNULA WERE USED. THE USER FACILITY STATES THAT THEY USED TO WORK WITH 80 TO 90% AMPLITUDE IN VASER MODE WITHOUT ANY PROBLEM. THEN THE PROBLEM OF BURNS STARTED TO OCCUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2242034 VASERLIPO SYSTEM SYSTEM, SUCTION, LIPOPLASTY QPB SOLTA MEDICAL, INC 110-0037 00850608002537
2865240 VASERLIPO SYSTEM SYSTEM, SUCTION, LIPOPLASTY QPB SOLTA MEDICAL, INC 110-0037 00850608002537

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention