RESPIRATORY HUMIDIFIER
Report
- Report Number
- 9611451-2009-00664
- Event Type
- Malfunction
- Date Received
- November 17, 2009
- Report Date
- November 20, 2009
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4) THE HC550 RESPIRATORY HUMIDIFIER IS NOT SOLD IN THE U.S., BUT IS SIMILAR TO A PRODUCT THAT IS SOLD IN THE U.S. THE 510(K) NUMBER OF THE SIMILAR PRODUCT IS K033710. METHOD: THE HC550 RESPIRATORY HUMIDIFIER AND ASSOCIATED (B)(4) TEMPERATURE PROBE WERE RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4). THE HUMIDIFIER WAS RUN AND TESTED IN ACCORDANCE WITH THE PRODUCT TECHNICAL MANUAL. THE ASSOCIATED TEMPERATURE PROBE WAS ALSO TESTED IN ACCORDANCE WITH THE TECHNICAL MANUAL. RESULTS: THE HUMIDIFIER AND TEMPERATURE PROBE FUNCTIONED PROPERLY AND PASSED ALL PERFORMANCE TESTS. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED HUMIDIFIER OR TEMPERATURE PROBE AS THE DEVICES FUNCTIONED PROPERLY DURING TESTING AND PASSED ALL PERFORMANCE TESTS.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT AN HC550 RESPIRATORY HUMIDIFIER OVERHEATS. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESPIRATORY HUMIDIFIER | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | HC550 | 061003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FISHER & PAYKEL HEALTHCARE (B)(4)| TEMPERATURE PROBE |