FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 15636624 · Received October 19, 2022

Report

Report Number
2647876-2022-00222
Event Type
Malfunction
Date Received
October 19, 2022
Date of Event
October 13, 2022
Report Date
February 15, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 2122277, MEDICAL DEVICE EXPIRATION DATE: 2023-02-28, DEVICE MANUFACTURE DATE: 2022-05-02. MEDICAL DEVICE LOT #: 2126025, MEDICAL DEVICE EXPIRATION DATE: 2023-02-28, DEVICE MANUFACTURE DATE: 2022-05-06. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: CATALOG: 442023. BATCH NO: 2122277. CUSTOMER REPORTED A PAENIBACILLUS CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BLOOD BACKGROUND WAS PERFORMED WITH SATISFACTORY RESULTS. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NONETHELESS A CORRECTIVE AND PREVENTIVE ACTION INITIATION DETERMINATION (CID) WAS PERFORMED TO EVALUATE THE NECESSITY OF A CAPA DUE TO THIS DEFECT. NO CAPA ACTION WAS REQUIRED. SEVERAL EVALUATIONS WERE PERFORMED TO CONCLUDE THAT THE BIO-CONTAMINANT REPORTED, PAENIBACILLUS, DID NOT ORIGINATED FROM THE BD MANUFACTURING SITE. THE AUTOCLAVE PROCESS WAS EVALUATED. THE MOIST HEAT AUTOCLAVE IS A PROCESS THAT USES MOIST HEAT AS THE LETHAL AGENT TO RENDER LIQUID AND POROUS/HARD GOODS ITEMS FREE FROM VIABLE MICROORGANISM AND RELIES ON DIRECT CONTACT WITH SATURATED STEAM, WHICH IS WATER VAPOR IN EQUILIBRIUM WITH LIQUID WATER. THE AUTOCLAVE PROCESS IS ACCOMPLISHED BY THE HEAT TRANSFER FROM THE STEAM TO THE LOAD AND BY HYDRATING EFFECT OF THE RESULTANT CONDENSATE THAT FORMS DUE TO THE CHANGE IN STATE FROM VAPOR TO THE LOWER ENERGY LIQUID STATE. THIS PHASE CHANGE REQUIRES THE TRANSFER OF THE LATENT HEAT OF THE STEAM TO THE SURROUNDINGS, THUS HEATING THE AUTOCLAVE AND ITS LOAD. BD CAYEY AUTOCLAVE THE PRODUCTS VIA MOIST HEAT. INITIAL VALIDATION HAD BEEN SUCCESSFULLY COMPLETED AND THE REQUIRED RE-QUALIFICATION ASSESSMENT HAS BEEN PERFORMED ACCORDINGLY MAINTAINING THE ACTIVE AND VALID AUTOCLAVING CYCLE. ISOLATES CONTAINING THE BIO-CONTAMINATION WERE RECEIVED FROM THE CUSTOMER. ENVIRONMENTAL SAMPLES WERE EVALUATED FROM BD MANUFACTURING SITE FOR IDENTIFICATION OF BACTERIAL SPECIES PRESENT AT DIFFERENT LOCATIONS WITHIN THE PLANT. THE PURPOSE OF IDENTIFYING THESE ISOLATES WAS TO DETERMINE IF PAENIBACILLUS CONTAMINATION ORIGINATED FROM THIS LOCATION. AFTER PERFORMING MALDI TESTING, THE CONTAMINANTS WERE IDENTIFIED AS PAENIBACILLUS SPECIES IN THE SAMPLES FROM THE CUSTOMER, AND NONE OF THE ENVIRONMENTAL SAMPLES COLLECTED FROM DIFFERENT LOCATIONS WITHIN THE MANUFACTURING PLANT IN BD CAYEY WERE IDENTIFIED AS PAENIBACILLUS. FINALLY, AUTOCLAVE TESTING OF PAENIBACILLUS SPECIES PRESENT IN ALL SAMPLES FROM THE CUSTOMER INDICATED THAT A 15-MINUTE AUTOCLAVE CYCLE AT 118°C IS SUFFICIENT TO KILL VEGETATIVE CELLS. BD CAYEY CYCLE IS SET FOR A HIGHER TEMPERATURE THAT EXCEEDS 15 MINUTES. THEREFORE, THE BIO-CONTAMINANT PAENIBACILLUS LIKELY DID NOT ORIGINATE FROM THE BD CAYEY, AND THERE IS NO EVIDENCE TO INDICATE THAT IT WOULD IT HAVE SURVIVED THE AUTOCLAVING PROCESS OF BACTEC BOTTLES. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. CATALOG: 442023. BATCH NO: 2126025. CUSTOMER REPORTED A PAENIBACILLUS CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BLOOD BACKGROUND WAS PERFORMED WITH SATISFACTORY RESULTS. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NONETHELESS A CORRECTIVE AND PREVENTIVE ACTION INITIATION DETERMINATION (CID) WAS PERFORMED TO EVALUATE THE NECESSITY OF A CAPA DUE TO THIS DEFECT. NO CAPA ACTION WAS REQUIRED. SEVERAL EVALUATIONS WERE PERFORMED TO CONCLUDE THAT THE BIO-CONTAMINANT REPORTED, PAENIBACILLUS, DID NOT ORIGINATED FROM THE BD MANUFACTURING SITE. THE AUTOCLAVE PROCESS WAS EVALUATED. THE MOIST HEAT AUTOCLAVE IS A PROCESS THAT USES MOIST HEAT AS THE LETHAL AGENT TO RENDER LIQUID AND POROUS/HARD GOODS ITEMS FREE FROM VIABLE MICROORGANISM AND RELIES ON DIRECT CONTACT WITH SATURATED STEAM, WHICH IS WATER VAPOR IN EQUILIBRIUM WITH LIQUID WATER. THE AUTOCLAVE PROCESS IS ACCOMPLISHED BY THE HEAT TRANSFER FROM THE STEAM TO THE LOAD AND BY HYDRATING EFFECT OF THE RESULTANT CONDENSATE THAT FORMS DUE TO THE CHANGE IN STATE FROM VAPOR TO THE LOWER ENERGY LIQUID STATE. THIS PHASE CHANGE REQUIRES THE TRANSFER OF THE LATENT HEAT OF THE STEAM TO THE SURROUNDINGS, THUS HEATING THE AUTOCLAVE AND ITS LOAD. BD CAYEY AUTOCLAVE THE PRODUCTS VIA MOIST HEAT. INITIAL VALIDATION HAD BEEN SUCCESSFULLY COMPLETED AND THE REQUIRED RE-QUALIFICATION ASSESSMENT HAS BEEN PERFORMED ACCORDINGLY MAINTAINING THE ACTIVE AND VALID AUTOCLAVING CYCLE. ISOLATES CONTAINING THE BIO-CONTAMINATION WERE RECEIVED FROM THE CUSTOMER. ENVIRONMENTAL SAMPLES WERE EVALUATED FROM BD MANUFACTURING SITE FOR IDENTIFICATION OF BACTERIAL SPECIES PRESENT AT DIFFERENT LOCATIONS WITHIN THE PLANT. THE PURPOSE OF IDENTIFYING THESE ISOLATES WAS TO DETERMINE IF PAENIBACILLUS CONTAMINATION ORIGINATED FROM THIS LOCATION. AFTER PERFORMING MALDI TESTING, THE CONTAMINANTS WERE IDENTIFIED AS PAENIBACILLUS SPECIES IN THE SAMPLES FROM THE CUSTOMER, AND NONE OF THE ENVIRONMENTAL SAMPLES COLLECTED FROM DIFFERENT LOCATIONS WITHIN THE MANUFACTURING PLANT IN BD CAYEY WERE IDENTIFIED AS PAENIBACILLUS. FINALLY, AUTOCLAVE TESTING OF PAENIBACILLUS SPECIES PRESENT IN ALL SAMPLES FROM THE CUSTOMER INDICATED THAT A 15-MINUTE AUTOCLAVE CYCLE AT 118°C IS SUFFICIENT TO KILL VEGETATIVE CELLS. BD CAYEY CYCLE IS SET FOR A HIGHER TEMPERATURE THAT EXCEEDS 15 MINUTES. THEREFORE, THE BIO-CONTAMINANT PAENIBACILLUS LIKELY DID NOT ORIGINATE FROM THE BD CAYEY, AND THERE IS NO EVIDENCE TO INDICATE THAT IT WOULD IT HAVE SURVIVED THE AUTOCLAVING PROCESS OF BACTEC BOTTLES. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) HAS CONTAMINATED BOTTLES DUE TO THE ISOLATION OF PAENIBACILLUS URINALIS GROUP. THE CUSTOMER PERFORMED GRAM STAIN AND ORGANISM ID USING BRUKER MALDI. CUSTOMER IS SEEING ABOUT ONE A DAY WITH ABOUT 60% STAINING AS GRAM NEGATIVE RODS AND THESE ARE THE ISOLATES THAT MIGHT CAUSE A PATIENT TO BE PUT ON UNNECESSARY ANTIBIOTICS. NO PATIENT IMPACT HAS BEEN REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER ALLEGES CONTAMINATED BOTTLES DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM SEVERAL BOTTLES. CUSTOMER PERFORMED GRAM STAIN AND ORGANISM ID USING BRUKER MALDI. BUT WE ARE SEEING ABOUT ONE A DAY NOW WITH ABOUT 60% STAINING AS GRAM NEGATIVE RODS AND THESE ARE THE ISOLATES THAT MIGHT CAUSE A PATIENT TO BE PUT ON UNNECESSARY ANTIBIOTICS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) HAS CONTAMINATED BOTTLES DUE TO THE ISOLATION OF PAENIBACILLUS URINALIS GROUP. THE CUSTOMER PERFORMED GRAM STAIN AND ORGANISM ID USING BRUKER MALDI. CUSTOMER IS SEEING ABOUT ONE A DAY WITH ABOUT 60% STAINING AS GRAM NEGATIVE RODS AND THESE ARE THE ISOLATES THAT MIGHT CAUSE A PATIENT TO BE PUT ON UNNECESSARY ANTIBIOTICS. NO PATIENT IMPACT HAS BEEN REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER ALLEGES CONTAMINATED BOTTLES DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM SEVERAL BOTTLES. CUSTOMER PERFORMED GRAM STAIN AND ORGANISM ID USING BRUKER MALDI. BUT WE ARE SEEING ABOUT ONE A DAY NOW WITH ABOUT 60% STAINING AS GRAM NEGATIVE RODS AND THESE ARE THE ISOLATES THAT MIGHT CAUSE A PATIENT TO BE PUT ON UNNECESSARY ANTIBIOTICS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2494525 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442023 SEE H.10 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 Unknown