FDA Adverse Event Malfunction Summary report: N

CHECK-IT

MDR report key: 15636133 · Received October 19, 2022

Report

Report Number
3016521623-2022-00444
Event Type
Malfunction
Date Received
October 19, 2022
Date of Event
October 7, 2022
Report Date
October 14, 2022
Manufacturer
LUCIRA HEALTH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO DEVICE(S) RETURNED. THEREFORE, A DHR REVIEW WAS COMPLETED FOR KIT LOT NUMBER K08A111803224M2 WITH 0 DISCREPANCIES FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THESE LOTS. THERE IS ONE SIMILAR COMPLAINT CMP-(B)(4) WITH AN ALLEGED FALSE POSITIVE REPORTED FROM THE SAME CUSTOMER ASSOCIATED WITH KIT LOT NUMBER K08A111803224M2. AN INVESTIGATION INTO THIS SIMILAR COMPLAINT HAS BEEN COMPLETED TO OBTAIN ADDITIONAL INFORMATION. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO (B)(4). (B)(4). (B)(4). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A111803224M2: SAMPLE VIAL LOT DHR REVIEW: 2202315, 2202461, 2202462, 2203051 ,2203099, 2203100 (ASSOCIATED INTERNAL LOT # 220124-1K, 220124-1Y, 220124-4F, 220124-4L, 220124-4Q, 220124-4T) TEST LOT DHRS REVIEW: 2203095, 2203133, 2203132, 2203133, 2203132, 2203135, 2203141, 2203142, 2203193, 2203192, 2203195, 2203197, 2202259 (ASSOCIATED INTERNAL LOT # 220305-3O, 220305-6J, 220305-6F, 220305-6I, 220305-6H, 220305-6G, 220305-6M, 220305-6Q, 220305-6Z, 220305-6W, 220305-9C, 220305-9K, 220124-15D). BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. WITH ANY POTENTIAL FALSE POSITIVE RESULT, THERE ARE SEVERAL POTENTIAL ROOT CAUSES: LOW VIRAL LOAD IN PATIENT SAMPLE THAT IS BELOW THE LOD OF THE LUCIRA AND REFERENCE TEST. AT VIRAL LOADS CLOSE TO LOD A TEST DETECTS A POSITIVE >95% OF THE TIME. WHEN THE VIRAL LOAD IS BELOW LOD, THE FOLLOW-UP TEST CANNOT PICK UP POSITIVE RELIABLY AND CAN GENERATE THE IMPRESSION OF A FALSE POSITIVE RESULT. LOW VIRAL LOADS CAN ALSO RESULT IN SAMPLING VARIABILITY BETWEEN SAMPLES. ENVIRONMENTAL CONTAMINATION FROM OTHER POSITIVES BEING TESTED IN THAT ENVIRONMENT OR DUE TO IMPROPER COLLECTION OR HANDLING OF THE SPECIMEN. DEVICE MALFUNCTION PLEASE NOTE OUR DEVICES ARE EXTENSIVELY INSPECTED AND TESTED THROUGH A ROBUST LOT RELEASE PROCESS AND THIS SCENARIO IS HIGHLY UNLIKELY. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA (B)(4) CHECK-IT.

Description of Event or Problem · 0

THE COMPLAINANT REPORTED AN ALLEGED FALSE POSITIVE RESULT. THE COMPLAINANT RETESTED WITH TWO ADDITIONAL LUCIRA TESTS WITH NEGATIVE RESULTS FOR CONFIRMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2656275 CHECK-IT COVID-19 TEST KIT QJR LUCIRA HEALTH, INC. K08A111803224M2

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other