FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 15629933 · Received October 19, 2022

Report

Report Number
9610877-2022-58976
Event Type
Malfunction
Date Received
October 19, 2022
Date of Event
September 27, 2022
Report Date
October 19, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL11-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K172156. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE LASER FILTER LIFTING. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE LIGHT GUIDE FIBER BUNDLE (LCB) BROKEN; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0587(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2938339 PENTAX VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1190STK

Patients

Seq Age Sex Outcome Treatment
1 Unknown