BD SYRINGE 0.5ML 31GA 6MM
Report
- Report Number
- 1920898-2022-00715
- Event Type
- Malfunction
- Date Received
- October 18, 2022
- Date of Event
- September 20, 2022
- Report Date
- November 11, 2022
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
INITIAL REPORTER NAME AND ADDRESS: INITIAL REPORTER PHONE #: (B)(6). DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. THE CUSTOMER RETURNED AN IMAGE OF A 0.5ML, 31 GUAGE, 6MM SYRINGE. THE CUSTOMER REPORTED THAT THE NEEDLE WAS STUCK INSIDE THE SYRINGE. THE PHOTOS WERE EXAMINED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS DETACHED FROM THE BARREL. NO DAMAGE TO THE BARREL TIP WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1172300. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE PHOTO RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATE. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. CORRECTIVE ACTION: CAPA HAS BEEN COMPLETED WHICH ADDRESSES THIS ISSUE. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT THE BD SYRINGE 0.5ML 31GA 6MM EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE STUCK INSIDE THE SYRINGE.
IT WAS REPORTED THAT THE BD SYRINGE 0.5ML 31GA 6MM EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE STUCK INSIDE THE SYRINGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2386451 | BD SYRINGE 0.5ML 31GA 6MM | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 1172300 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |