FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 15626659 · Received October 18, 2022

Report

Report Number
1911916-2022-00567
Event Type
Malfunction
Date Received
October 18, 2022
Date of Event
September 21, 2022
Report Date
September 23, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028307
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. IT WAS REPORTED SYRINGES WERE EITHER MISSING THE HUB SCALE MARKINGS PARTIALLY OR ENTIRELY. TO AID IN THE INVESTIGATION, SIX SAMPLES IN SEALED PACKAGING BLISTERS AND THREE PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THE SYRINGES HAVE SCALE MARKING ISSUES. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302830, LOT NUMBER 2152001. THE REVIEW REVEALED THERE WAS AN INCIDENT REPORTED DURING THE PRODUCTION OF THIS BATCH. IT COULD BE POSSIBLE THE PRODUCTS RECEIVED WERE ESCAPES FROM THIS INCIDENT. THE PRODUCTS IDENTIFIED AS AFFECTED WERE SEGREGATED AND SCRAPPED. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD LUER-LOK¿ TIP SYRINGES EXPERIENCED MISSING SCALE MARKINGS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE SEVERAL BD 20 ML SYRINGES EITHER MISSING THE HUB DEMARCATIONS ENTIRELY OR PARTIALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340719 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 302830 2152001 30382903028307

Patients

Seq Age Sex Outcome Treatment
1 Unknown