FDA Adverse Event Other Summary report: N

BRAINSCAN

MDR report key: 1562527 · Received December 17, 2009

Report

Report Number
8043933-2009-00012
Event Type
Other
Date Received
December 17, 2009
Report Date
December 4, 2009
Manufacturer
BRAINLAB AG
Product Code
MUJ
PMA / PMN Number
K994413
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE NUMBER OF PTS THAT MIGHT HAVE BEEN TREATED WITH A LOWER THAN INTENDED RADIATION DOSE AT THIS HOSPITAL, AND THE POTENTIAL CLINICAL EFFECTS HAVE NOT BEEN REVEALED BY THE HOSPITAL TO BRAINLAB, THUS IS UNK TO BRAINLAB AT THIS POINT OF TIME. ACCORDING TO THE HOSPITAL, THE BRAINSCAN TREATMENT PLANNING SOFTWARE HAS BEEN USED WITH THE ERRONEOUS DOSE DATA SINCE 2005. ERRONEOUS DOSE DATA IMPLEMENTATION BY HOSPITAL POTENTIALLY RESULTING IN UNINTENDED PT TREATMENT. THE BRAINLAB INSTRUCTIONS FOR THE ACCORDING DOSE DATA ACQUISITION BY THE HOSPITAL WERE CORRECT AND COMPLETE. THERE WAS NO QUALITY OR PERFORMANCE ISSUE OR MALFUNCTION OF BRAINLAB EQUIPMENT. CORRECTIVE AND PREVENTIVE ACTIONS: SINCE THERE WAS NO QUALITY OR PERFORMANCE ISSUE OR MALFUNCTION OF BRAINLAB EQUIPMENT, THERE ARE NO REMEDIAL ACTIONS INTENDED BY BRAINLAB AT THIS POINT OF TIME.

Description of Event or Problem · 1

THE HOSPITAL STATES, THAT INCORRECT DOSE DATA REGARDING THE LINEAR ACCELERATOR CALIBRATION WAS ENTERED INTO THE BRAINLAB TREATMENT PLANNING SOFTWARE AT THE TIME OF THE TREATMENT PLANNING COMMISSIONING. AS A RESULT, AN APPROXIMATELY 15% LOWER THAN INTENDED RADIATION DOSE MIGHT HAVE BEEN DELIVERED TO PTS SINCE (B)(6) 2005 AT THIS HOSPITAL. BRAINLAB BRAINSCAN TREATMENT PLANNING SOFTWARE VERSION 5.31 WAS INVOLVED. THE INITIAL DOSE DATA ACQUISITION, ALL SUBSEQUENT MAINTENANCE AND QUALITY ASSURANCE OF THE DOSE DATA IMPLEMENTED INTO THE BRAINLAB TREATMENT PLANNING SOFTWARE IS IN THE RESPONSIBILITY OF THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BRAINSCAN RADIOTHERAPY TREATMENT PLANNING SW MUJ BRAINLAB AG BRAINSCAN V. 5.31 NA

Patients

Seq Age Sex Outcome Treatment
1