FDA Adverse Event Injury Summary report: N

MECTACER HEAD BIOLOX OPTION Ø 28

MDR report key: 15624382 · Received October 18, 2022

Report

Report Number
3005180920-2022-00795
Event Type
Injury
Date Received
October 18, 2022
Date of Event
September 30, 2022
Report Date
October 18, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809323
PMA / PMN Number
K131518
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 18-OCT-2022: LOT 22C0009: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01-MAR-2022. EXPIRATION DATE: 2027-02-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICES INVOLVED: BATCH REVIEWS PERFORMED ON 18-OCT-2022: BALL HEADS: MECTACER 01.29.242A SLEEVE SIZE L (K131518) LOT 2114384: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-JAN-2022. EXPIRATION DATE: 2027-01-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. LINER: VERSAFITCUP DM 01.26.2854MHC DOUBLE MOBILITY HC LINER Ø 54/28 (K092265) LOT 2202287: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-APR-2022. EXPIRATION DATE: 2027-03-13. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2351747 MECTACER HEAD BIOLOX OPTION Ø 28 HIP CERAMIC HEAD LZO MEDACTA INTERNATIONAL SA 01.29.230H 22C0009 07630030809323

Patients

Seq Age Sex Outcome Treatment
1 Female Required Intervention