IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM
Report
- Report Number
- 0002023141-2022-02611
- Event Type
- Injury
- Date Received
- October 17, 2022
- Date of Event
- September 16, 2022
- Report Date
- February 24, 2023
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019508
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). PATIENT WEIGHT NOT PROVIDED. ADDITIONAL 510(K) NUMBERS ARE K011028 AND K013227.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ONE (1) IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM (TSVB10) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED BONE DEBRIS ON EXTERNAL IMPLANT THREADS. THE IMPLANT EXHIBITS A CONCAVE HAIRLINE FRACTURE IN THE COLLAR AREA. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. A PRE-EXISTING CONDITION NOTED ON THE PER WAS LOW BONE DENSITY (TYPE III). THE REPORTED IMPLANT WAS LOCATED ON TOOTH # 35 (FDI) AND WAS USED FOR APPROXIMATELY 5 MONTHS, 15 DAYS. THE LENGTH OF THE SCREW USAGE IS UNKNOWN. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1251791). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1251791) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE IMPLANT. & FRACTURE SCREW BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR, AND THE REPORTED EVENTS WERE CONFIRMED.
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BECOMING FRACTURED AT THE SITE. THE PROCEDURE WAS ABLE TO BE COMPLETED USING ANOTHER IMPLANT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2224885 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVB10 | 1251791 | 00889024019508 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Female | Required Intervention |