FDA Adverse Event Injury Summary report: N

UNK MAXERA 58MM ACETABULUM

MDR report key: 15614722 · Received October 17, 2022

Report

Report Number
0001822565-2022-02975
Event Type
Injury
Date Received
October 17, 2022
Report Date
October 17, 2022
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
NI
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). REPORT SOURCE CANADA. DOI: HTTPS://DOI.ORG/10.1186/S12891-022-05082-6. REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED IN A JOURNAL ARTICLE APPROXIMATELY 44 MONTHS POST LEFT TOTAL HIP ARTHROPLASTY THE PATIENT EXPERIENCED A LOUD NOISE ON THE LEFT SIDE AND FELT A SHOCK AND ACUTE PAIN IN THE LEFT LIMB. RADIOGRAPHIC EVIDENCE DISPLAYED PNEUMARTHROSIS LATERALLY TO THE ACETABULUM, WHICH REABSORBED OVER A TWO MONTH SPAN. TWO MONTHS AFTER THE INCIDENT, THE PATIENT UNDERWENT A REVISION AND FOUND THE CERAMIC LINER SLIGHTLY DISPLACED IN THE METAL CUP BUT REMAINED STABLE. TAPER MODULAR JUNCTION WEAR AND CORROSION SIGNS WERE OBSERVED. THE MAXERA CUP WAS REMOVED AND REPLACED WITH A 2MM LARGER CUP. NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2944251 UNK MAXERA 58MM ACETABULUM PROSTHESIS, HIP LPH ZIMMER BIOMET, INC. N/A NI

Patients

Seq Age Sex Outcome Treatment
1 49 YR Male Hospitalization| R UNK BIOLOX DELTA FEMORAL HEAD 48MM| UNK BIOLOX DELTA TAPER LINER| UNK TITANIUM SHELL