BD PHASEAL¿ SPIKE CONNECTOR
Report
- Report Number
- 3003152976-2022-00466
- Event Type
- Malfunction
- Date Received
- October 15, 2022
- Date of Event
- September 21, 2022
- Report Date
- December 1, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- ONB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY: ONE SAMPLE WAS RETURNED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PRODUCT WAS VISUALLY INSPECTED AND THE CONNECTOR WAS OBSERVED TO BE DISCONNECTED FROM THE SPIKE. THERE WAS TRACES OF SOLVENT OBSERVED ON THE C35 CONNECTOR PIECE, HOWEVER, IT WAS NOTED TO BE AN INSUFFICIENT AMOUNT, RESULTING IN IMPROPER ASSEMBLY WHICH CAUSED THE PIECES TO DISCONNECT. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2201210, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. PRODUCT UNDERGOES INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, NO ISSUES WERE FOUND. ALTHOUGH WE COULD NOT IDENTIFY A DIRECT ISSUE, IT WAS DETERMINED THIS INCIDENT WAS LIKELY RELATED TO A FAILURE IN THE DISPENSER THAT DOSES THE SOLVENT, RESULTING IN THE IMPROPER ASSEMBLY OF THE CONNECTOR ONTO THE SPIKE. A PROJECT HAS BEEN INITIATED TO FURTHER INVESTIGATE AND PREVENT ANY REOCCURRENCE OF THIS ISSUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND THE REPORTED CONDITION WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.
IT WAS REPORTED THAT 30 BD PHASEAL¿ SPIKE CONNECTORS EXPERIENCED SPONTANEOUS DISCONNECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE BONDING OF THE SPIKE PART AND THE CONNECTOR WAS SEPARATED (THESE TWO PARTS HAVE NOT BEEN BONDED).
IT WAS REPORTED THAT 30 BD PHASEAL¿ SPIKE CONNECTORS EXPERIENCED SPONTANEOUS DISCONNECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE BONDING OF THE SPIKE PART AND THE CONNECTOR WAS SEPARATED (THESE TWO PARTS HAVE NOT BEEN BONDED).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2857279 | BD PHASEAL¿ SPIKE CONNECTOR | INTRAVASCULAR ADMINISTRATION SET | ONB | BECTON DICKINSON, S.A. | 2201210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |