FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE

MDR report key: 15611794 · Received October 14, 2022

Report

Report Number
1213809-2022-00717
Event Type
Malfunction
Date Received
October 14, 2022
Date of Event
September 20, 2022
Report Date
September 21, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903029953
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT#: 2070214, MEDICAL DEVICE EXPIRATION DATE: 28-FEB-2027, DEVICE MANUFACTURE DATE: 11-MAR-2022, MEDICAL DEVICE LOT#: UNKNOWN, MEDICAL DEVICE EXPIRATION DATE: UNKNOWN, DEVICE MANUFACTURE DATE: UNKNOWN. INVESTIGATION SUMMARY: THREE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE BARREL CRACK WAS OBSERVED EXTENDING FROM THE 1.0ML TO THE 4.0ML GRAD LINES. POTENTIAL ROOT CAUSE FOR THE BARREL CRACKED DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. THESE CONDITIONS ARE OCCURRING BELOW THEIR EXPECTED FREQUENCY SO NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER: 2070214. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD LUER-LOK¿ SYRINGES EXPERIENCED A CASE OF DEVICE DAMAGE WHILE STILL CONSIDERED OPERABLE, AND A FAILURE TO CONTAIN BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT HAPPENED AGAIN THIS MORNING. THE BARREL IS CRACKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2876896 BD LUER-LOK¿ SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 302995 SEE H10 30382903029953

Patients

Seq Age Sex Outcome Treatment
1 Unknown