FDA Adverse Event Summary report: N

KLEENSPEC

MDR report key: 1559326 · Received December 9, 2009

Report

Report Number
1559326
Date Received
December 9, 2009
Date of Event
November 11, 2009
Report Date
December 9, 2009
Manufacturer
WELCH ALLYN, INC.
Product Code
FTF
Report Source
User Facility report
Reporter Location
CT, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

NURSE ATTEMPTED TO START A PERIPHERAL INTRAVENOUS CATHETER USING A TRANSILLUMINATOR FOR ASSISTANCE LOCATING A VEIN. INSTEAD OF THE TRANSILLUMINATOR, HE MISTAKENLY USED A VAGINAL ILLUMINATOR, WHICH CAUSED THE PATIENT TO RECEIVE 2ND DEGREE BURNS ON THE PALM AND ON THE FOREARM. THE VAGINAL ILLUMINATOR LOOKS SIGNIFICANTLY SIMILAR TO THE TRANSILLUMINATOR AND IT HAD BEEN PLACED IN THE SAME AREA WHERE THE TRANSILLUMINATOR WAS KEPT. ALTHOUGH THE NURSE RECOGNIZED IT DID NOT LOOK EXACTLY THE SAME, HE KNEW THAT THERE WERE PLANS TO PURCHASE NEW TRANSILLUMINATORS RECENTLY AND MISTAKENLY ASSUMED THE VAGINAL ILLUMINATOR WAS ONE OF THE NEW ONES. IT WAS NOT.THESE TWO PIECES OF EQUIPMENT ARE NEVER STORED TOGETHER AND WE DO NOT KNOW HOW IT GOT WHERE IT WAS. WE HOPE TO GET TO THE ROOT CAUSE OF THAT, AS WELL AS FORMULATE ACTION PLANS SO THAT THIS DOES NOT HAPPEN AGAIN.

Description of Event or Problem · 1

NURSE ATTEMPTED TO START A PERIPHERAL INTRAVENOUS CATHETER USING A TRANSILLUMINATOR FOR ASSISTANCE LOCATING A VEIN. INSTEAD OF THE TRANSILLUMINATOR, HE MISTAKENLY USED A VAGINAL ILLUMINATOR, WHICH CAUSED THE PATIENT TO RECEIVE 2ND DEGREE BURNS ON THE PALM AND ON THE FOREARM. THE VAGINAL ILLUMINATOR LOOKS SIGNIFICANTLY SIMILAR TO THE TRANSILLUMINATOR AND IT HAD BEEN PLACED IN THE SAME AREA WHERE THE TRANSILLUMINATOR WAS KEPT. ALTHOUGH THE NURSE RECOGNIZED IT DID NOT LOOK EXACTLY THE SAME, HE KNEW THAT THERE WERE PLANS TO PURCHASE NEW TRANSILLUMINATORS RECENTLY AND MISTAKENLY ASSUMED THE VAGINAL ILLUMINATOR WAS ONE OF THE NEW ONES. IT WAS NOT.THESE TWO PIECES OF EQUIPMENT ARE NEVER STORED TOGETHER AND WE DO NOT KNOW HOW IT GOT WHERE IT WAS. WE HOPE TO GET TO THE ROOT CAUSE OF THAT, AS WELL AS FORMULATE ACTION PLANS SO THAT THIS DOES NOT HAPPEN AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KLEENSPEC ILLUMINATOR, VAGINAL FTF WELCH ALLYN, INC. 78810 *

Patients

Seq Age Sex Outcome Treatment
1 12 DA NO OTHER THERAPIES| NO OTHER THERAPIES