FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME
MDR report key: 1559263
·
Received December 4, 2009
Report
- Report Number
- 1824206-2009-07033
- Event Type
- Malfunction
- Date Received
- December 4, 2009
- Date of Event
- November 30, 2009
- Report Date
- November 30, 2009
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
TECH FOUND HEAD UP SOLENOID VALVE WAS BAD. TECH REPLACED HEAD UP SOLENOID VALVE TO RESOLVE ISSUE.
Description of Event or Problem · 1
HEAD WORKS INTERMITTENTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM RITTER | P1900C001566 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |