FDA Adverse Event
Malfunction
Summary report: N
OMNIPOD INSULIN PUMP
MDR report key: 15591764
·
Received October 12, 2022
Report
- Report Number
- 3004464228-2022-19073
- Event Type
- Malfunction
- Date Received
- October 12, 2022
- Date of Event
- September 16, 2022
- Report Date
- October 6, 2022
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K192659
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO CONFIRM THE REPORTED NEEDLE MECHANISM FAILURE OR DETERMINE ITS ROOT CAUSE. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.
Additional Manufacturer Narrative · 0
CORRECTION TO D(4): CATALOG NO CHANGED FROM UNAVAILABLE TO ZXP425. CORRECTION TO G(5): PMA/510(K) # CHANGED FROM UNAVAILABLE TO K192659.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE NEEDLE DEPLOYED EARLY. POD WAS NOT WORN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2575105 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 40160 | L50296 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female |