FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 15585052
·
Received October 12, 2022
Report
- Report Number
- 9610877-2022-58694
- Event Type
- Malfunction
- Date Received
- October 12, 2022
- Date of Event
- September 20, 2022
- Report Date
- October 12, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- UDI-DI
- 04961333234066
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC34-I10TL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K180285. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE IMAGE CLOUDY (NOT CLEAR VIEW). BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE MOISTURE CONDENSATION ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY ).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2665370 | PENTAX | VIDEO COLONOSCOPE | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC34-I10TL | 04961333234066 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |